Installation Instructions
Aseptoflex adapter
E330xx / E331xx
701306/04 05/06
UK
UK
UK
UK
UK
2
Mount
adapter (C) to sensor (A) first,
then sensor + adapter to the pro-
cess connection by means of a nut,
a clamping flange or similar (B).
If it is not possible to slide the fixing
element (B) down over the top of the
sensor: slide it up over the bottom
of the sensor before the adapter is
mounted.
A
B
C
NOTE: A guarantee for a long-term stable and maintenance-free fitting, with no
bug traps in the hygienic sealing of the metal seal (Aseptoflex connection) is only
valid for once-only mounting.
Mounting of the adapter:
Grease thread and sealing chamfer 1.
of the sensor and the contact face
and of the adapter with the lubrica-
ting paste supplied.
The lubricating paste is food-grade
(USDA-H1 84-201).
Make sure that the O-ring (D) is
correctly positioned.
D
1
1 = Grease
Screw the sensor into the adapter 2.
until it is hand-tight. Do not damage
the sealing chamfers.
I
3
UK
Clamp sensor and adapter into a 3.
clamping device (E). The sealing
chamfers (F) must not be damaged.
Tighten the sensor slightly using a
spanner until you can feel the end
stop (corresponding to a maxi-
mum tightening torque of 25 Nm /
18 ft-lb).
E
F
Note: Do not overtighten. This can have adverse effect on the sealing.