Installation instructions
Electronic pressure sensor
for industrial applications
PT24xx
80230780 / 00 07 / 2017
UK
2
1 Preliminary note
Symbols used
Instructions
Cross-reference
Important note
Non-compliance can result in malfunction or interference.
Information
Supplementary note.
2 Safety instructions
Please read this document prior to set-up of the unit. Ensure that the product is
suitable for your application without any restrictions.
If the operating instructions or the technical data are not adhered to, personal
injury and/or damage to property can occur.
Improper or non-intended use may lead to malfunctions of the unit or to
unwanted effects in your application. That is why installation, electrical
connection, set-up, operation and maintenance of the unit must be carried out
by qualified personnel authorised by the machine operator.
In order to guarantee the correct condition of the unit for the operating time it
is necessary to use the unit only for media to which the wetted materials are
sufficiently resistant (Technical data).
Contents
1 Preliminary note 2 ...................................................................................................
2 Safety instructions 2 ...............................................................................................
3 Functions and features 3 ........................................................................................
3.1 Applications ...................................................................................................3
4 Functions .............................................................................................................4
5 Installation............................................................................................................4
6 Electrical connection 4 ............................................................................................
7 Technical data and scale drawing 5 ........................................................................
3
UK
The responsibility whether the measurement units are suitable for the
respective application lies with the operator. The manufacturer assumes no
liability for consequences of misuse by the operator. Improper installation and
use of the units result in a loss of the warranty claims.
3 Functions and features
The pressure sensor detects the system pressure and converts it into an analogue
output signal.
3.1 Applications
Type of pressure: relative pressure
Order number
Final value of the
measuring range
(nominal pressure)
Static pressure
resistance (max.
permissible pressure)
Bursting
pressure
psi psi psi
PT2415 100 250 2900
PT2424 200 580 6525
PT2434 300 940 8700
PT2443 500 1450 11600
PT2402 1000 2500 13050
PT2432 3000 7250 14500
PT2400 5000 14500 24650
Avoid overload pressure exceeding the specified maximum permissible
pressure by taking appropriate measures.
The indicated bursting pressure must not be exceeded.
Even if the bursting pressure is exceeded only for a short time, the unit
may be destroyed.
ATTENTION: Risk of injury!
Pressure Equipment Directive (PED):
The units comply with the Pressure Equipment Directive and are designed
and manufactured for group 2 fluids in accordance with the sound
engineering practice.
Use of group 1 fluids on request!
The units are vacuum resistant.