Tripp Lite Isolator IS300HGDV Manual de Usario

Tripp Lite No categorizado Isolator IS300HGDV

Lee a continuación 📖 el manual en español para Tripp Lite Isolator IS300HGDV (24 páginas) en la categoría No categorizado. Esta guía fue útil para 8 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios

Página 1/24
Owner’s Manual
Isolator® Series 115/230V
Medical-Grade™
Isolation Transformers
with Step-Up and Step-Down Capabilities
IS300HGDV
(Series Number AG-8798)
IS600HGDV
(Series Number AG-8799)
IS1000HGDV
(Series Number AG-879A)
IS1800HGDV
(Series Number AG-879B)
1111 W. 35th Street, Chicago, IL 60609 USA • www.tripplite.com/support
Copyright © 2017 Tripp Lite. All rights reserved. All trademarks are the sole property of their respective owners.
Table of Contents
1. Overview 2
2. Important Safety Instructions 2
3. Installation 3
4. Optional Connections 5
5. Troubleshooting 5
6. Maintenance 6
7. Warranty 6
Español 7
Français 13
Русский 19
2
1. Overview
2. Important Safety Instructions
Statement of Intended Use
Tripp Lite 115/230V ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. Listed Medical-Grade Isolation Transformers offer line isolation,
step-up and step-down capabilities and continuous noise filtering. An internal low-impedance isolation transformer component offers 100%
isolation from the input AC line. Full ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. medical-grade listing makes it ideal for protection of
medical electronic equipment in patient-care vicinities. Faraday shield and floating AC output reduces the cumulative leakage current of
the Isolator and connected equipment to levels less than 100 microamps. Transformer filtering provides continuous common-mode noise
rejection with no wearable parts. All units are equipped with a potential equalization bolt (POAG) and ground lug for connection to a medical
potential equalization installed in the room. Rugged all-metal unit housing; ships ready for placement in any industrial, medical, office or
network environment. Step-up and step-down transformer (115/230V 50/60Hz) with universal IEC-320-C14 or IEC-320-C20 inputs and
IEC-320-C13 outputs allows for worldwide use.
The above features also make these products ideal for audio-video isolation, as well as other similar applications.
Transformer Location Warnings
Install the transformer indoors, away from excess moisture or heat, conductive contaminants, dust or direct sunlight.
For best performance, keep the indoor temperature at 77°F (25°C). The operational range is between 32° and 104°F (0° and 40°C).
Transformer Connection Warnings
Connect the transformer directly to a properly grounded AC power outlet.
Do not modify the transformer’s plug, and do not use an adapter that would eliminate the transformer’s ground connection.
To remove AC voltage from the transformer, pull the plug from the socket outlet.
Remove mains supply before installing, uninstalling, fuse replacement or system changes.
Observe accident prevention and safety regulations that apply specially to the application.
Before start-up check whether the rated voltage of the equipment matches the output voltage selection on the transformer.
Do not use this equipment within oxygen-enriched atmospheres or within 1 ft. (0.3 m) at a point from which an oxygen-enriched
atmosphere is intentionally ventilated.
Equipment Connection Warnings
Use of this equipment in life support applications where failure of this equipment can reasonably be expected to cause the failure of the
life support equipment or to significantly affect its safety or effectiveness is not recommended.
To reduce the risk of fire, connect only to a circuit that has branch circuit over current protection with an ampere rating in accordance with
the National Electrical Code® (NEC®), ANSI/NFPA 70 or your local electrical code. In Europe, the circuit breaker must meet the IEC/EN
60934 standard and have a contact air gap of at least 3 mm.
3
1
2
3
1
2
3
Use a small, flat blade screwdriver to open
the access cover and fuse holder for the
fuses and to set the output voltage.
Use a small, flat blade screwdriver to open
the access cover and fuse holder for the
fuses and to set the output voltage.
3. Installation
Before connecting equipment to the isolation transformer, verify the local mains voltage matches the input voltage
selected on the transformer. The output voltage must match the voltage compatible with the equipment load.
The appliance inlet of the isolation transformer (300VA, 600VA or 1000VA) is designed so power cannot be applied
while the fuse access door is opened. If this protective feature is damaged, have the appliance inlet replaced by a
qualified service person. Do not defeat this protective feature or attempt to perform inspection or replacement of
fuses while the isolation transformer is ON and plugged into mains power.
Install in accordance with applicable electrical codes, such as National Electrical Code and/or NFPA 99 Article 517. Use only with an
approved plug/cord set according to the equipment electrical ratings being used. The transformer must be installed by a qualied technician
only. Compliance with ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. must be evaluated after the final installation.
3.1 Select input and output voltage on the isolation transformer based on the local mains
voltage available and the voltage rating of the equipment to be connected.
3.1.1 For IS300HGDV and IS600HGDV:
Input voltage selection is made via the inlet fuse holder (Figure 1). The voltage selected will display above the power switch A. You
will also have to insert the correct fuse value depending on the transformer model and input voltage. For correct fuse value, see the
table in section 3.1.5. Output voltage selection is made with the selector switch B.
If using as a step-up or step-down transformer, the output voltage must match the connected equipment load requirements.Note:
Figure 1
Figure 2
A
A
B
B
Output Voltage
Selector Switch
Output
Voltage
Selector
Switch
3.1.2 For IS1000HGDV:
Input voltage selection is made via the inlet fuse holder (Figure 2). The voltage selected will display above the power switch A. You
will also have to insert the correct fuse value depending on input voltage. For correct fuse value, see the table in section 3.1.5.
Output voltage selection is made with the selector switch B.
Note: If using as a step-up or step-down transformer, the output voltage must match the connected equipment load requirements.

Especificaciones del producto

Marca: Tripp Lite
Categoría: No categorizado
Modelo: Isolator IS300HGDV
Color del producto: Wit
Peso.: 1800 g
Ancho: 217.9 mm
Profundidad: 284.9 mm
Altura: 53.3 mm
Código de seguridad internacional (IP): IP68
Indicadores LED: Ja
Actual: 0.8 A
Colocación: Muur
Banda de frecuencia: 2.4, 5 GHz
Cables suministrados: LAN (RJ-45)
Incluye adaptador de CA: Ja
Número de puertos Ethernet LAN (RJ-45): 2
Consumo de energía (máx.): 35.7 W
Certificación: FCC, IC
Conexión para adaptador de red: Ja
Memoria interna: 256 MB
Modulación: 64-QAM, BPSK, CCK, DBPSK, DQPSK, OFDM, QPSK
LAN Ethernet, velocidades de transferencia de datos: 10,100,1000 Mbit/s
Estándar de red: IEEE 802.11a, IEEE 802.11ac, IEEE 802.11b, IEEE 802.11d, IEEE 802.11g, IEEE 802.11n, IEEE 802.1Q, IEEE 802.1x, IEEE 802.3at
Algoritmos de seguridad compatibles: 802.1x RADIUS, AES, SSID, TKIP, WPA, WPA-EAP, WPA-PSK, WPA2
Cantidad de antenas: 6
Nivel de ganancia de antena (máx.): 7 dBi
Número de usuarios: 100 gebruiker(s)
Temperatura de almacenamiento: -30 - 80 °C
Interno: Nee
Es posible actualizar el firmware: Ja
Velocidad de transferencia de datos de LAN inalámbrica (máx.): 1300 Mbit/s
Gestión basada en web: Ja
Humedad de almacenamiento: 0 - 90 procent
Botón de reinicio: Ja
Filtrado de direcciones MAC: Ja
filtro de dirección IP: Ja
Memoria flash: 16 MB
Servidor DHCP: Ja
Velocidad máxima de transferencia de datos: 1300 Mbit/s
2,4GHz: Ja
5GHz: Ja
Canal de Banda ancha: 40 MHz
Calidad de servicio (QoS): Ja
VLAN=soporte: Ja
Wi-Fi multimedia (WMM)/(WME): Ja
Funciones de LAN virtual: Tagged VLAN
cliente DHCP: Ja
Alimentación a través de Ethernet (PoE): Ja
Funciones de antena: Afneembare antenne
Limitación de velocidad: Ja
Tipo de interruptor: Managed
Número de canales: 11 kanalen
puerto WAN: Ethernet (RJ-45)
Protocolos de gestión: SNMP v1/v2c, MIB I/II
transmitir potencia: 15 dBmW
voltaje-de-entrada-de-CA: 100 - 240 V
Frecuencia-de-entrada-de-CA: 50 - 60 Hz
Temperatura de funcionamiento (TT): -20 - 70 °C
Humedad relativa de funcionamiento (VV): 0 - 90 procent
Código del Sistema Armonizado (SA): 85176990
Número de soportes SSID: 16

¿Necesitas ayuda?

Si necesitas ayuda con Tripp Lite Isolator IS300HGDV haz una pregunta a continuación y otros usuarios te responderán




No categorizado Tripp Lite Manuales

No categorizado Manuales

Últimos No categorizado Manuales