SBI-505 Indicator
USER MANUAL
Rev. A
Safety Instruction
For safety operation please follow the safety instruction.
WARNING
Settings, Calibration, Inspection and Maintenance should only be
done by a trained person or staff.
WARNING
Please make sure the indicator has sufficient grounding.
WARNING
The indicator is a static sensitive piece of
equipment. Cut off the power during electrical
connections and internal components touched by
hand are prohibited.
Please take all measure of anti-static protection.
1
Instruction
This indicator is designed for basic weighing function.
Main function
Weighing function:
Zero, tare, G.W, N.W, accumulation, printing, animal -weighing.
kg/lb conversion.
Print format: S.N. G.W N.W Tare. Date, Time
Options:
Pinter
RS232/RS485 serial interface or second display
Technical parameter
Accuracy class 6000 e
Resolution display: 30, 000 ADC: 2,000,000
Zero stability error TK0 < 0.1μV//K
Span stability error TKspn < ± 6 ppm//K
Sensitivity (internal) 0. 3 μV /d
Input voltage -30~30mV DC
Excitation circuit 5 VDC, 4 wire connection,
Maximum connect 6 load cell of 350
AC power AC100~250V
Operation temperature - 10 °C ~ + 40 °C
Operation humidity 90%RH
Storage temperature - 40 °C ~ + 70 °C

Especificaciones del producto

Marca: Brecknell
Categoría: Libra
Modelo: SBI-505