Omron CompAIR Pro NE-C900 Manual de Usario
Lee a continuaciรณn ๐ el manual en espaรฑol para Omron CompAIR Pro NE-C900 (4 pรกginas) en la categorรญa Inhalador. Esta guรญa fue รบtil para 2 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios
Pรกgina 1/4
EN
Intended use
Medical Purpose
This product is intended to be used for inhaling medication for respiratory
disorders.
Intended User
๎๎Legally certified medical experts, such as doctor, nurse and therapist.
๎๎Caregiver or patient under the guidance of qualified medical experts for
home treatment.
๎๎The user should also be capable of understanding general operation of
NE-C900 and the content of instruction manual.
Intended Patients
This product should not be used on patients, who are unconscious or are
not breathing spontaneously.
Environment
This product is intended for use in a medical facility, such as Hospital,
clinic and doctorโs office, and in a room of general household.
Durable period
Durable periods are as follows, provided the product is used to nebulize
2 ml of medication 10 times a day for 6 minutes each time at room
temperature (23ยฐC).
Durable period may vary depending on usage environment.
Compressor (Main unit) 5 years
Nebulizer Kit 1 year
Mouthpiece 1 year
Nosepiece 1 year
Air Tube (PVC, 200cm) 1 year
Air Filter 60 days
Mask (PVC)/(SEBS) 1 year
Precautions for use
Warnings and cautions described in the instruction manual should be
observed.
Know your unit
Important safety instructions
Read all the information in the instruction manual and any other literature included in
the box before using the device.
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in
serious injury.
(Usage)
๎๎ For type, dose, and regime of medication follow the instructions of your doctor or
respiratory therapist.
๎๎ If you feel anything unusual during use, stop using the device immediately and
consult your doctor.
๎๎ Do not use only water in the nebulizer for inhaling purposes.
๎๎ Keep the device out of the reach of unsupervised infants and children. The device
may contain small pieces that can be swallowed.
๎๎ Do not store the air tube while there is moisture or medication remaining inside.
๎๎ Do not use or store the device where it may be exposed to noxious fumes or volatile
substances.
๎๎ Do not use the device where it may be exposed to flammable gas.
๎๎ Do not cover the compressor with a blanket, towel, or any other type of cover during
use.
๎๎ Always dispose of any remaining medication after use, and use fresh medication each
time.
๎๎ Do not use in anaesthetic or ventilator breathing circuits.
๎๎ Do not block the ventilation slots. Never place the device where the ventilation slots
may be obstructed during operation.
๎๎ Store the parts in a clean location for avoiding infection.
(Risk of electrical shock)
๎๎ Never plug or unplug the device with wet hands.
๎๎ The compressor and power cord are not waterproof. Do not spill water or other liquids
on these parts. If liquid does spill on these parts, immediately unplug the
power cord
and wipe off the liquid with gauze or other absorbent material.
๎๎ Do not immerse the compressor in water or other liquid.
๎๎ Do not use or store the device in humid locations.
๎๎ Do not operate the device with a damaged power cord or plug.
๎๎ Keep the power cord away from heated surfaces.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these
rules may result in damage, inefficient nebulization or infection. For the instructions,
refer to โCleaning and disinfectingโ section.
๎๎ Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using
them when as follows:
- the first time after purchase.
- if the device has not been used for a long period of time
- if more than one person uses the same device
๎๎ Be sure to wash or wipe the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
๎๎ Do not leave the cleaning solution in the parts.
Caution: Indicates a potentially hazardous situation which if not avoided, may result in
minor or moderate injury, or physical damage.
(Usage)
๎๎ Provide close supervision when this device is used by, on, or near children or invalids.
๎๎ Make sure that the parts are attached correctly.
๎๎ Make sure that the air filter is correctly attached.
๎๎ Make sure that the air filter is clean. If the air filter has been used for more than 60
days, replace it with a new one.
๎๎ Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or
shake it while in use.
๎๎ Do not use or store the device while the air tube is creased.
๎๎ Use only original nebulizing parts, air tube, air filter and air filter cover.
๎๎ Do not add more than 7 ml of medication to the medication tank.
๎๎ Do not carry or leave the nebulizer kit while the medication tank contains medication.
๎๎ Do not leave the device unattended with infants or persons who cannot express their
consent.
๎๎ Do not subject the device or the parts to any strong shocks such as dropping the device
on the floor.
๎๎ Do not jab the nozzle of the medication tank with a pin or any sharp object.
๎๎ Do not insert fingers or objects inside the compressor.
๎๎ Do not disassemble or attempt to repair the compressor or power cord.
๎๎ Do not leave the device or its parts in extreme hot or cold temperature, or under direct
sunlight.
๎๎ Do not block the air filter cover.
๎๎ Do not use the device while sleeping or if drowsy.
๎๎ Approved for human use only.
๎๎ When using the device, the main unit may become hot.
๎๎ Do not touch the main unit for other than necessary operation such as turning off the
power while nebulizing.
๎๎ To avoid the medication residue on the face, be sure to wipe the face after removing
the mask.
๎๎ To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of
the nose.
๎๎ Do not use a damaged nebulizer kit, mouthpiece or nosepiece.
(Risk of electrical shock)
๎๎ Always unplug the device from the power outlet after use and before cleaning.
๎๎ Plug the device into the appropriate voltage outlet. Do not overload power outlets or
use extension cords.
๎๎ Do not misuse the power cord.
๎๎ Do not wind the power cord around the compressor.
๎๎ Changes or modifications not approved by OMRON HEALTHCARE will void the
user warranty.
๎๎ Do not pull the power cord strongly.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these
rules may result in damage, inefficient nebulization or infection. For the instructions,
refer to โCleaning and disinfectingโ section.
๎๎ Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
๎๎ Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
๎๎ When disinfecting parts by boiling, make sure that the container does not boil dry.
Otherwise it may also result in fire.
Accessories
Name Model
Nebulizer Kit & Mouthpiece Set NEB-NSET3-90E
Adult Mask (PVC) NEB-MSLP-E
Child Mask (PVC) NEB-MSMP-E
Infant Mask (PVC) NEB-MSSP-E
Adult Mask (SEBS) NEB-MSLS-E
Child Mask (SEBS) Set NEB-MSSS-E
Mouthpiece NEB-MP-E
Nosepiece NEB-NP-E
Air Tube (PVC, 200cm) NEB-TP-L20E
Air Filters x 5 NEB-AFR-28E
Thank you for purchasing the OMRON COMP AIR Pro.
This product was developed in conjunction with respiratory therapists
for the successful treatment of asthma, chronic bronchitis, allergies
and other respiratory disorders.
This is a medical device. Operate the device only as instructed by
your doctor and/or respiratory therapist.
How the nebulizer kit works
The medication that is pumped up
through the medication channel is
mixed with compressed air which is
generated by a compressor pump.
The compressed air mixed with
medication is turned into fine particles
and sprayed when in contact with the
baffle.
Manufacturer OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com
Production facility OMRON (DALIAN) CO., LTD.
Dalian, CHINA
Subsidiary
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY
www.omron-healthcare.de
OMRON SANTร FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Made in China
General Safety Precautions:
๎๎ Inspect the device and parts before using them each time, and check that there are
no problems. In particular, be sure to check the following:
- That the nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
๎๎ When using this device, there will be some noise and vibration caused by the pump in
the compressor. There will also be some noise caused by the emission of compressed
air from the nebulizer kit. This is normal and does not indicate a malfunction.
๎๎ Operate the device only as intended. Do not use the device for any other purpose.
๎๎ Do not use the device at temperatures greater than +40ยฐC.
๎๎ Make sure that the air tube is securely attached to the compressor and nebulizer kit,
and does not come loose. Twist the air tube plug slightly when inserting it into the air
tube connectors to avoid the air tube disconnecting during use.
๎๎ To completely isolate the device from the power source unplug the plug from the
power source.
Keep these instructions for future reference.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PCโs and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic interference
from other devices. Electromagnetic interference may result in incorrect operation of the
medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with
the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has
been implemented. This standard defines the levels of immunity to electromagnetic
interferences as well as maximum levels of electromagnetic emissions for medical
devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
๎๎ Do not use mobile (cellular) telephones and other devices, which generate strong
electrical or electromagnetic fields, near the medical device. This may result in
incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of
the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be
disposed of, with other household wastes at the end of its working life. To prevent
possible harm to the environment or human health from uncontrolled waste disposal,
please separate this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product,
or their local government office, for details of where and how they can return this item
for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the
purchase contract. This product should not be mixed with other commercial wastes for
disposal.
TM
Baffle
Aerosol
Medication
Channel
Medication
Aerosol
Nozzle
Medication
Rear View
Power Cord
Ventilation Slots
Power Plug
Ventilation Slots
Power Switch
Air Filter Cover
*Air filter inside
Air Tube
Connector
Nebulizer Kit Holder
Inhalation Top
Medication Tank
Compressor Nebulizer
Model NE-C900
Instruction Manual
Compressed Air
Compressor (Main Unit) x 1
Nebulizer Kit x 1 Nosepiece x 1Mouthpiece x 1
Adult Mask (PVC) x 1 Child Mask (PVC) x 1
Air Tube
(PVC, 200cm) x 1
Instruction Manual
Warranty card
Spare Air Filters x 5
IM-NE-C900-E-EN-04-08/2013
1675345-3D
Technical data
Name: COMP AIR Pro
Model: NE-C900 (NE-C900-E)
Type: Compressor Nebulizer
Rating: AC 220-240V, 50Hz
Power Consumption: 220VA
Operating Temperature/
Humidity:
+10ยฐC to +40ยฐC / 30% to 85%RH
Storage and Transport
Temperature/Humidity/
Air Pressure:
-20ยฐC to +60ยฐC / 10% to 95%RH /
700 - 1060hPa
Weight: Approx. 2.2kg (compressor only)
Dimensions: Approx. 175 (W) ร 110 (H) ร215 (D) mm
(compressor only)
Contents:
Compressor, Nebulizer Kit, Air Tube (PVC,
200cm), Mouthpiece, Nosepiece, Adult Mask
(PVC), Child Mask (PVC), 5 pcs Spare Air
Filters, Instruction Manual, Warranty Card.
Classification: Class ll equipment, Type B applied part
= Class ll
equipment
= Type B
applied part
Read the instruction
manual carefully
= Power off = Power on
Technical data for the OMRON NE-C900 Compressor with the Nebulizer Kit:
Particle Size: **MMAD Approx. 3ฮผm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity: 7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Sound: *Approx. 55dB
Nebulization Rate: *Approx. 0.4ml/min (by weight loss)
Airflow/Compressor
Output:
*7L/min @100kPa
Aerosol Output: **0.57ml (2ml, 1%NaF)
Aerosol Output Rate: **0.08ml/min (2ml, 1%NaF)
Result of cascade impactor measurements for particle size**
* Measured by OMRON HEALTHCARE Co., Ltd.
**
Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009.
Notes:
๎๎ Subject to technical modification without prior notice.
๎๎ This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan.
๎๎ The device may not work if the temperature and voltage conditions are different to
those defined in the specifications.
๎๎ This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory
therapy equipment - Part1: Nebulizing systems and their components.
๎๎ Performance may vary with drugs such as suspensions or high viscosity. See drug
supplierโs data sheet for further details.
๎๎ See website of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the
compressor.
2) Remove the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.
Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new
one.
1. Pull the air filter cover off the compressor.
2. Change the air filter.
Remove the old air filter with a sharp object
and insert the new air filter.
3. Put the air filter cover back in place.
Make sure it is attached properly.
Notes:
๎๎ Use only OMRON air filters designed for this device. Do not operate without a filter.
๎๎ Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters
can cause blockages.
๎๎ There is no front or back orientation for the air filters.
๎๎ Check that air filters are clean and free of dust before inserting them.
Cleaning and disinfecting
Cleaning
Clean the parts after each use to remove residual medication. This will avoid
inefficient nebulization or infection.
Q Washable parts
๎๎Nebulizer Kit (Inhalation top, Medication tank), Mask (PVC)/(SEBS),
Mouthpiece, Nosepiece
Wash them in warm water and mild detergent (neutral detergent). Rinse
them thoroughly with clean hot tap water and allow to air dry in a clean
place.
Q Non-washable parts
๎๎Compressor, Air Tube
Firstly, make sure that the power plug is unplugged from the power outlet.
Wipe clean with a soft cloth moistened with water or mild detergent
(neutral detergent).
๎๎Air Filter
Do not wash or clean the air filter. If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
Disinfecting
Always disinfect the parts after the last treatment of the day. If the parts are
heavily stained, replace them with new ones.
To select a method for disinfection, refer to the table below.
o: applicable X: not applicable
Disinfecting
method
Parts
Mouthpiece
Nosepiece
Nebulizer Kit
Adult Mask (SEBS)
(optional)
Child Mask (SEBS)
(optional)
Adult Mask (PVC)
Child Mask (PVC)
Infant Mask (PVC)
(optional)
Air Tube
(PVC, 200cm)
Air Filter
Materials
PP
PP
PP
Mask: SEBS
Band: Rubber
(non-latex)
Adapter: PP
Mask: PVC
(Phthalate free)
Band: Rubber
(non-latex)
PVC
(Phthalate free)
Polyester
A
Alcohol
o o o o o X X
Disinfecting
ethanol
Sodium
hypochlorite o o o o o X X
Milton*
(0.1%, 15min.)
Quaternary
ammonium o o o o o X X
Osvan*
(0.1%, 10min.)
Chlorhexidine
o o o o o X X
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant o o o o o X X
Tego 51*
(0.2%, 15min.)
BBoiling o o o o X X X
*๎ an example of commercially available disinfectant. (The concentration and residence
time specified in the table are under the conditions where the service life of the parts
is tested with each disinfectant used as described in its instruction manual. Please
note that the testing was not performed for the purpose of ensuring the effectiveness
of the disinfectants. There is no intention of suggesting that such disinfectants be
used. Conditions of use and ingredients of disinfectant vary for manufacturers.
Please read the instruction manuals carefully before use and apply disinfectant to
each part in an appropriate way. Please notice that service life of the parts may be
shorter depending on conditions, environments and frequency of use.)
A. Use a commercially available disinfectant. Follow the instructions
provided by the disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake
off excess water and allow to air dry in a clean
environment.
How to use
Warning:
Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using them
when as follows:
- the first time after purchase.
- if the device has not been used for a long period of time
- if more than one person uses the same device
For the instructions, refer to โCleaning and disinfectingโ section.
1. Make sure that the power switch is in the off ( ) position.
2. Plug the power plug into a power outlet.
3. Remove the inhalation top from the medication
tank.
4. Add the correct amount of prescribed medication
to the medication tank.
5. Put the inhalation top back inside the medication
tank.
6. Attach the mask, mouthpiece or nosepiece to the nebulizer kit tightly.
7. Attach the air tube.
Twist the air tube plug slightly and push it firmly into the air tube connector.
8. Hold the nebulizer kit as
indicated on the right.
Follow the instructions of
your doctor or respiratory
therapist.
Caution:
Do not tilt the nebulizer kit at an
angle of greater than 45 degrees
in all directions. Medication may
flow into the mouth or it may result
in not nebulizing effectively.
9. Press the power switch to the on ( ) position. As the compressor
starts, nebulization begins and aerosol is generated.
Inhale the medication.
10. When treatment is complete, turn the power off and unplug the
compressor from the power outlet.
Troubleshooting
Check the following if your unit should fail during operation.
Problem Cause Remedy
Nothing happens when the
power switch is pressed.
Is the power plug plugged into
a power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication tank.
Is there too much/little
medication in the medication
tank?
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked? Make sure that the nozzle is free
of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit is
not tilted at an angle of more than
45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked? Make sure that the air tube is free
of blockages.
Is the air filter dirty? Replace the air filter with a new
one.
The device is abnormally
loud.
Is the air filter cover attached
correctly?
Attach the air filter cover correctly.
The device is very hot. Is the compressor covered?
Do not cover the compressor with
any type of cover during use.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device
- no parts of the unit are user serviceable. Return the unit to an authorized
OMRON retail outlet or distributor.
1
2
1
2
Clicks
Right angle
45ยฐ45ยฐ
100
90
80
70
60
50
40
30
20
10
0
1 10 1000.1
Cumulative % particle mass of sodium fluoride under size
Cumulative Undersize%
๎๎
๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎๎
Individual tests
Mean
Especificaciones del producto
Marca: | Omron |
Categorรญa: | Inhalador |
Modelo: | CompAIR Pro NE-C900 |
ยฟNecesitas ayuda?
Si necesitas ayuda con Omron CompAIR Pro NE-C900 haz una pregunta a continuaciรณn y otros usuarios te responderรกn
Inhalador Omron Manuales
14 Septiembre 2024
12 Septiembre 2024
12 Septiembre 2024
15 Julio 2024
15 Julio 2024
15 Julio 2024
Inhalador Manuales
- Inhalador Beurer
- Inhalador Medisana
- Inhalador Rossmax
- Inhalador Microlife
- Inhalador Scala
- Inhalador Drive Medical
- Inhalador PARI
รltimos Inhalador Manuales
12 Octubre 2024
7 Octubre 2024
23 Septiembre 2024
16 Septiembre 2024
15 Septiembre 2024
14 Septiembre 2024
14 Septiembre 2024
13 Septiembre 2024
28 Agosto 2024
14 Agosto 2024