Earth testers
EN - User’s manual
C.A 6422
C.A 6424
2
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
WARNING, risk of electric shock. The voltage applied to parts marked with this symbol may be hazardous.
Useful information or tip. Earth.
Battery. Ammeter clamp.
The product is declared recyclable following a life cycle analysis in accordance with standard ISO 14040.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete life-
cycle has enabled us to control and optimize the e󰀨ects of the product on the environment. In particular this appliance
exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates compliance with the European Low Voltage Directive (2014/35/EU), the Electromagnetic
Compatibility Directive (2014/30/EU), the Directive on Radioelectric Equipment (2014/53/EU) and the Directive on the
Restriction of Hazardous Substances (RoHS, 2011/65/EU and 2015/863/EU).
The UKCA marking certies that the product is compliant with the requirements that apply in the United Kingdom, in
particular as regards Low-Voltage Safety, Electromagnetic Compatibility, and the Restriction of Hazardous Substances.
The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective disposal
in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.
Denition of measurement categories
Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
Example: power feeders, counters and protection devices.
Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or xed industrial devices.
Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations.
Example: power supply to electro-domestic devices and portable tools.
PRECAUTIONS FOR USE
This instrument is compliant with safety standard IEC IEC/EN 61010-2-30 or BS EN 61010-2-30 for voltages up to 600V in category IV.
Do not use the instrument for measurements on circuits that are not in measurement categories II, III, or IV or that might be con-
nected inadvertently to circuits that are not in measurement categories II, III, or IV.
The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken
in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
If you use this instrument other than as specied, the protection it provides may be compromised, thereby endangering you.
Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
Before each use, check the condition of the insulation on the leads, housing, and accessories. Any item of which the insulation
is deteriorated (even partially) must be set aside for repair or scrapping.
Before using your instrument, check that it is perfectly dry. If it is wet, it must be thoroughly dried before it can be connected or
used.
Use only the leads and accessories supplied. The use of leads (or accessories) of a lower voltage or category limits the voltage
or category of the combined instrument and leads (or accessories) to that of the leads (or accessories).
Use personal protection equipment systematically.
Thank you for purchasing a C.A 6422 or C.A 6424 earth tester.
To obtain the best service from your instrument:
read this user manual carefully,
comply with the precautions for use.
3
CONTENTS
1. FIRST START-UP .................................................................................................................................................................4
1.1. Delivery condition 4 ........................................................................................................................................................
1.2. Accessories .................................................................................................................................................................4
1.3. Inserting the batteries 4 ..................................................................................................................................................
1.4. Primary batteries in the C.A 6424 or rechargeable batteries in the C.A 6422 5 .............................................................
1.5. Charging the battery (C.A 6424) ..................................................................................................................................6
1.6. Carrying the device 7 ......................................................................................................................................................
1.7. Use on a desktop 7 .........................................................................................................................................................
2. PRESENTATION OF THE DEVICES ..................................................................................................................................8
2.1. C.A 6422 ......................................................................................................................................................................8
2.2. C.A 6424 ......................................................................................................................................................................9
2.3. Functions of the instruments 10 ....................................................................................................................................
2.4. Keys and buttons 10 .......................................................................................................................................................
2.5. Display unit ................................................................................................................................................................ 11
3. USE ......................................................................................................................................................................................12
3.1. Voltage measurement (C.A 6424) 12 .............................................................................................................................
3.2. Resistance measurement (2P) 13 ..................................................................................................................................
3.3. Earth resistance measurement 15 ..................................................................................................................................
3.4. Current measurement (C.A 6424) .............................................................................................................................19
4. TECHNICAL CHARACTERISTICS .................................................................................................................................. 20
4.1. General reference conditions 20 ....................................................................................................................................
4.2. Electrical characteristics 20 ............................................................................................................................................
4.3. Variations in the range of use 22 ....................................................................................................................................
4.4. Intrinsic uncertainty and operating uncertainty 23 ..........................................................................................................
4.5. Environmental conditions 23 ..........................................................................................................................................
4.6. Power supply 23 .............................................................................................................................................................
4.7. Mechanical characteristics 24 .......................................................................................................................................
4.8. Conformity to international standards 24 ........................................................................................................................
4.9. Electromagnetic compatibility (CEM) 24 .........................................................................................................................
5. MAINTENANCE ..................................................................................................................................................................25
5.1. Cleaning 25 ....................................................................................................................................................................
5.2. Replacing the batteries 25 ..............................................................................................................................................
6. WARRANTY ........................................................................................................................................................................26
When handling the leads, test probes, and crocodile clips, keep your ngers behind the physical guard.
All troubleshooting and metrological checks must be performed by competent and accredited personnel.