Megohmmeters
EN - User’s manual
C.A 6532
C.A 6534
2
Thank you for purchasing a megohmmeter C.A 6532 or C.A 6534.
For best results from your instrument:
read these operating instructions carefully,
comply with the precautions for use.
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
WARNING, risk of electric shock. The voltage applied to parts marked with this symbol may be hazardous.
Equipment protected by double insulation. Earth.
The voltage on the terminals must not exceed 700 V. Battery.
Remote control probe. Information or useful tip.
The product is declared recyclable following an analysis of the life cycle in accordance with standard ISO14040.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete
lifecycle has enabled us to control and optimize the e󰀨ects of the product on the environment. In particular this appliance
exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates compliance with the European Low Voltage Directive (2014/35/EU), Electromagnetic
Compatibility Directive (2014/30/EU), Radio Equipment Directive (2014/53/EU), and Restriction of Hazardous Substances
Directive (RoHS, 2011/65/EU and 2015/863/EU).
The UKCA marking certies that the product is compliant with the requirements that apply in the United Kingdom, in
particular as regards Low-Voltage Safety, Electromagnetic Compatibility, and the Restriction of Hazardous Substances.
The rubbish bin with lines through it indicates that, in the European Union, the product must undergo selective disposal
in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.
Denition of measurement categories
Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
Example: power feeders, counters and protection devices.
Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or xed industrial devices
Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations.
Example: power supply to electro-domestic devices and portable tools.
> 700V
3
PRECAUTIONS FOR USE
This instrument is compliant with safety standard IEC/EN 61010-2-034 or BS EN 61010-2-034 and the leads are compliant with
IEC/EN 61010-031 or BS EN 61010-031, for voltages up to 600 V in category IV or 1,000 V in category III.

installations.
The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in
use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.

The safety of any system in which this instrument might be incorporated is the responsibility of the integrator of the system.
This instrument can be used on category IV installations, for voltages not exceeding 600 V with respect to earth or 700 V 
max between terminals.
Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
Observe the environmental conditions of use.
Except for voltage measurements, make no measurements on live devices.
Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
Before each use, check the condition of the insulation on the leads, housing, and accessories. Any item of which the insulation
is deteriorated (even partially) must be set aside for repair or scrapping. There is a risk of electric shock if the instrument is used
without its battery compartment cover.
Before using your instrument, check that it is perfectly dry. If it is wet, it must be thoroughly dried before it can be connected or used.
Use only the leads and accessories supplied. The use of leads (or accessories) of a lower voltage rating or category limits the use
of the combined instrument + leads (or accessories) to the lowest category and service voltage.

Before removing of the battery compartment cover, make sure that the measurement leads (and accessories) are disconnected.
Replace all of the batteries at once. Use alkaline batteries.
Use personal protection equipment systematically.
All troubleshooting and metrological checks must be done by competent, accredited personnel.