Omron BP742N Manual de Usario
Omron
monitor de presión arterial
BP742N
Lee a continuación 📖 el manual en español para Omron BP742N (2 páginas) en la categoría monitor de presión arterial. Esta guía fue útil para 13 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios
Página 1/2
Introduction
Thank you for purchasing the OMRON® BP742N Blood Pressure Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and
pulse rate in adult patient population who can understand this instruction manual
with the arm circumference range printed on the arm cuff. The device detects the
appearance of irregular heartbeats during measurement and gives a warning signal
with the measurement result.
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR PHYSICIAN.
Important Safety Information
Warning: Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Consult your physician before using the device in pregnancy including
pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood
transfusion.
Consult your physician before using the device on the arm with an arterio-venous
(A-V) shunt.
Do not use the device with other medical electrical (ME) equipment
simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner
exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation in
infants.
Contained small parts that may cause a choking hazard if swallowed by infants.
(AC Adapter (optional) Usage)
Do not use the AC adapter if the device or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a
multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Caution: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
People with severe blood flow problems, or blood disorders, should consult a
physician before using the device, as the arm cuff inflation can cause bruising.
Remove the arm cuff if it does not start deflating during the measurement.
Do not use this device on infants or persons who cannot express their intentions.
Do not use the device for any purpose other than measuring blood pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs may result
in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields,
near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff. This may cause an inaccurate
reading.
Do not use in a location with moisture, or a location where water may splash on
the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Do not take measurements more than necessary. It may cause bruising due to
blood flow interference.
Consult your physician before using the device if you had a mastectomy.
Read "If your systolic pressure is more than 210 mmHg" of this instruction
manual, if your systolic pressure is known to be more than 210 mmHg. Inflating
to a higher pressure than necessary may result in bruising where the arm cuff is
applied.
(AC Adapter (optional) Usage)
Fully insert the power plug into the outlet.
When disconnecting the power plug from the outlet, do not pull the power cord.
Be sure to pull from the power plug safely.
When handling the power cord, take care not to do the following:
Wipe the dust off from the power plug.
Unplug monitor when not in use.
Disconnect the power plug before cleaning.
Use only an OMRON AC adapter designed for this device. Use of unsupported
adapters may damage and/or may be hazardous to the device.
(Battery Usage)
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 “AA” alkaline or manganese batteries with this device. Do not use
other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
General Precautions
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the monitor, not
the tube itself.
• Do not drop the monitor or subject device to strong shocks or vibrations.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
inaccurate reading.
• Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental pollution.
1. Know Your Device
Contents:
Monitor, arm cuff, battery set, instruction manual, quick start guide
Monitor:
Arm cuff:
Display:
1.1 Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular
rhythm two or more times during the
measurement, the irregular heartbeat
symbol ( ) will appear on the display
with the measurement values.
An irregular heartbeat rhythm is defined
as a rhythm that is 25% less or 25%
more than the average rhythm detected
while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results,
we recommend you consult your physician. Follow the directions of your
physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement, remain still during measurement.
Average Value Symbol ( )
The average value symbol is displayed when you press and hold the memory
button for more than 3 seconds. The most recent average value appears on
the display screen.
Blood Pressure Level Indicator
2013 ESH/ESC Guidelines for the management of arterial hypertension
Definitions of hypertension by office and home blood pressure levels
These are from statistical values for blood pressure.
1.2 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating
for 30 minutes before taking a measurement. Rest for at least 15 minutes
before taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful
times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so
that the arm cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement does not provide an accurate indication of
your true blood pressure. You need to take and record several readings
over a period of time. Try to measure your blood pressure at the same time
each day for consistency.
2. Preparation
2.1 Battery Installation
1. Remove the battery cover.
2. Insert 4 “AA” batteries as
indicated in the battery
compartment.
3. Replace the battery cover.
Notes:
• When the low battery symbol ( ) appears on the display, turn the monitor
off, then replace all batteries at the same time. Long life alkaline batteries
are recommended.
• The measurement values continue to be stored in memory even after the
batteries are replaced.
• The supplied batteries may have a shorter life.
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
2.2 Setting the Date and Time
Set the monitor to the correct date and time before taking a measurement for
the first time.
1. Press the button.
2. Push or to change the year.
Push button to confirm the year and then the month flashes. Repeat the
same steps to change the month, day, hour, and minute.
3. Press the START/STOP button to turn the monitor off.
Notes:
• If the batteries have been replaced, the date and time setting will need to be
reset.
• If the date and time are not set, “-:--” appears during or after measurement.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.
1. Insert the air plug into the air jack
securely.
2. Wrap the arm cuff firmly in
place around your left upper
arm.
The bottom edge of the arm cuff
should be 1/2 inch (1 to 2 cm) above
the elbow. Air tube is on the inside of
your arm and aligned with your middle
finger.
3. Secure closed with the
fabric fastener.
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on the
air tube.
• The blood pressure can differ between the right arm and the left arm, and
the measured blood pressure values can be different. OMRON recommends
to always use the same arm for measurement. If the values between both
arms differ substantially, please check with your physician which arm to use
for your measurements.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. Avoid
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at
the same level as your heart.
3.3 Taking a Measurement
Notes:
• To stop a measurement, press the START/STOP button once to deflate the
arm cuff.
• Remain still while taking a measurement.
The monitor is designed to take measurements and store the measurement
values in the memory for 2 people using USER ID 1 and USER ID 2.
1. Select your User ID (1 or 2).
2. Press the START/STOP button.
The arm cuff will start to inflate automatically.
3. Remove the arm cuff.
4.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement result in its memory.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another measurement. Waiting between
measurements allows the arteries to return to the condition prior to
taking a measurement.
Using the Guest Mode
The monitor stores measurement values for 2 users in the memory.
The guest mode can be used to take a single measurement for another user.
No measurement values are stored in the memory when the guest mode is
selected.
1. Press and hold the START/STOP
button for more than 3 seconds.
The USER ID symbol and the Date/Time
display will disappear.
2. Release the START/STOP button when the Date/Time display
turns off.
The arm cuff will start to inflate automatically.
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
Inflating to a higher pressure than necessary may result in bruising where the
arm cuff is applied.
3.4 Using the Memory Function
The monitor automatically stores the results up to 50 sets for each user (1 and
2). It can also calculate an average value based on the last 3 measurement
values taken within 10 minutes.
Notes:
• If there are only 2 measurement values in the memory for that period, the
average will be based on these 2 values.
• If there is 1 measurement value in the memory for that period, this is
displayed as the average.
• If the memory is full, the monitor will delete the oldest value.
• When viewing the measurement value taken without setting the date and
time, “-:--” is displayed instead of the date and time.
To View the Measurement Values Stored in Memory
1. Select your User ID (1 or 2).
2. Press the button.
The Memory number appears for a second
before the pulse rate is displayed.
The newest set is numbered “1”.
3. Press the or button to view the values stored in memory.
: To view the older values
: To view the more recent values
To View the Average Value
1. Select your User ID (1 or 2).
2. Press and hold the button for
more than 3 seconds.
Notes:
• If the previous measurement was taken without setting the date and time,
the average value is not calculated.
• If there are no measurements results stored in the memory,
the screen to the right is displayed.
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by USER ID.
1. Select your User ID (1 or 2).
2. Press the button, while the memory symbol appears.
3. While holding the button
down, press the
START/STOP button for
more than 3 seconds.
Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.
Do not damage. Do not break it.
Do not tamper with it. Do not forcibly bend or pull.
Do not twist. Do not bundle during use.
Do not pinch. Do not place under heavy objects.
2298802-0B
5 series
Blood Pressure Monitor
Model BP742N
Instruction Manual
A. Display
B. Up/Down buttons
C. USER ID selection switch
D. Date/Time setting button
E. START/STOP button
F. Memory button
G. Battery compartment
H. Air jack
I. AC adapter jack
(for optional AC adapter)
M. USER ID symbol
N. Systolic blood pressure
O. Diastolic blood pressure
P. Low battery symbol
Q. Heartbeat symbol
(Flashes during measurement.)
R. Memory symbol
S. Average value symbol
T. Date/Time display
U. Movement error symbol
V. Irregular heartbeat symbol
W. Blood pressure level indicator
X. Deflation symbol
Y. Pulse display / Memory number
Office Home
Systolic Blood Pressure ≥ 140 mmHg 135 mmHg≥
Diastolic Blood Pressure ≥ 90 mmHg ≥ 85 mmHg
G
H
I
D
A
F
E
B
C
D
D
J
L
K
J. Arm cuff
(Arm circumference 9” - 17” (22 - 42 cm))
K. Air plug
L. Air tube
N
O
P
R
M
S
T
Q
U
V
W
X
Y
Irregular Heartbeat
Normal Heartbeat
Pulse
Blood pressure
Blood pressure
Short Long
Pulse
Minute
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Year
Month
Day
Hour
: Forward
: Back
1/2 inch
(1 - 2 cm)
START INFLATE DEFLATE COMPLETE
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Alternating date/time
display
4. Error Messages and Troubleshooting
4.1 Error Messages
4.2 Troubleshooting
5. Maintenance and Storage
5.1 Maintenance
To protect your device from damage, please observe the following:
• Store the device and the components in a clean, safe
location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse
them in water.
• Do not use gasoline, thinners or similar solvents to clean the
device.
• Use a soft and dry cloth, or a soft and moistened cloth and
neutral soap to clean on the monitor and the arm cuff.
• Changes or modification not approved by the manufacturer
will void the user warranty. Do not disassemble or attempt to
repair the device or components. Consult Customer Service.
5.2 Storage
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm cuff.
Note: Do not bend or crease the air tube
excessively.
Do not store the device in the following situations:
•If the device is wet.
•Locations exposed to extreme temperatures, humidity, direct sunlight, dust or
corrosive vapors such as bleach.
•Locations exposed to vibrations, shocks or where it will be at an angle.
5.3 Optional Medical Accessories
Using the Optional AC Adapter
1. Insert the AC adapter plug
into the AC adapter jack on
the rear side of the monitor.
2. Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet
first, and then remove the AC adapter plug from the monitor.
6. Specifications
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for
determination of diastolic blood pressure.
• This device has not been validated for use on pregnant patients.
7. FCC Statement
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
8. Limited Warranty
Your BP742N Automatic Blood Pressure Monitor, excluding the arm cuff and
batteries, is warranted to be free from defects in materials and workmanship
appearing within 2 years from the date of purchase, when used in accordance with
the instructions provided with the monitor. The arm cuff is warranted to be free from
defects in materials and workmanship appearing within 1 year from the date of
purchase when the monitor is used in accordance with the instructions provided
with the monitor. The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor or arm cuff covered by the
above warranty. Replacement is our only responsibility and your only remedy under
the above warranty.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for
the address of the inspection center and the return shipping and handling fee.
Enclose the original printed receipt. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to
prevent damage in transit. Because of possible loss in transit, we recommend
insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF
THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL,
INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR
DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
9. Guidance and Manufacturer’s Declaration
OMRON Automatic Blood Pressure Monitor
Information for accompanying documents in the scope of IEC60601-1-2:2007
Model: BP742N including AC-adapter
Error Display Cause Solution
Irregular heartbeats are
detected.
Remove the arm cuff. Wait 2 - 3
minutes and then take another
measurement. Repeat the steps
in section 3.3. If this error
continues to appear, contact
your physician.
Movement during measurement. Carefully read and repeat the
steps in section 3.3.
The batteries are low.
You should replace the batteries
with new ones ahead of time.
Refer to section 2.1.
The batteries are exhausted.
You should replace the batteries
with new ones at once.
Refer to section 2.1.
Air plug disconnected. Insert the plug securely.
Refer to section 3.1.
Arm cuff is applied too loosely. Apply the arm cuff tighter.
Refer to section 3.1.
Air is leaking from the arm cuff.
Replace the arm cuff with a new
one.
Refer to section 5.3.
Movement during measurement
and the arm cuff has not been
inflated sufficiently.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the arm cuff manually
until it is 30 to 40 mmHg above
your previous measurement
result.
Refer to section 3.3.
The arm cuff was inflated
exceeding the maximum
allowable pressure, and then
deflated automatically when
inflating the arm cuff manually.
Do not touch the arm cuff and/or
bend the air tube while taking a
measurement. Do not inflate the
arm cuff more than necessary.
Refer to section 3.3.
Movement during measurement.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.1.
Device error. Contact Customer Service.
Problem Cause and Solution
No power.
No display appears on the monitor.
Replace all batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Refer to section 2.1.
Measurement values appear too high or
too low.
Blood pressure varies constantly. Many
factors including stress, time of day, and
how you wrap the cuff, may affect your
blood pressure. Review the section 1.2
and section 3.3.
Arm cuff AC Adapter
Arm circumference
9” - 17” (22 - 42 cm)
CD-WR17
(Model: HEM-RML31)
HEM-ADPTW5
Model BP742N (HEM-7131-Z)
Display LCD digital display
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy Pressure: 3 mmHg or 2% of reading
Pulse: 5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Measurement method Oscillometric method
IP classification IP 20
Power source
4 “AA” batteries 1.5V or optional AC adapter (INPUT AC100-240V
50/60Hz 0.12A)
Battery life Approximately 1000 measurements (using new alkaline batteries)
Operating temperature
/ humidity 50°F to 104°F (10°C to 40°C) / 15 to 90% RH
Storage temperature /
humidity / air pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060hPa
Weight Monitor : Approximately 10 5/8 oz. (300g) not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Dimensions Monitor : Approximately 4 1/4” (w) 3 1/8” (h) 5 1/2” (l)
(107mm 79mm 141mm)
Arm cuff : Approximately 5 3/4” 23 1/2” (air tube: 29 1/2”)
(145 mm 594 mm (air tube: 750 mm))
Cuff circumference 9” to 17” (220 to 420 mm)
Memory Up to 50 per user
Contents
Monitor, arm cuff, battery set, instruction manual, quick start guide
FOR CUSTOMER SERVICE
Visit our web site at: www.omronhealthcare.com
Call toll free: 1-800-634-4350
= Type BF
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic interference from
other devices. Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation. Medical devices should also not interfere with
other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to
prevent unsafe product situations, the IEC60601-1-2 standard has been implemented. This
standard defines the levels of immunity to electromagnetic interferences as well as maximum
levels of electromagnetic emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this IEC60601-1-2:2007
standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specified by OMRON, with the exception
of cables sold by OMRON as replacement parts for internal components, may result in
increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other equipment.
In case adjacent or stacked use is necessary, the medical device should be observed to
verify normal operation in the configuration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device should
be used.
• The MEDICAL ELECTRICAL EQUIPMENT BP742N including AC-adapter needs special
precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in this documentations.
• The Essential Performance of the BP742N including AC-adapter is to measure a blood
pressure and a pulse rate and using the memory function.
The BP742N including AC-adapter may be interfered with by other equipment, even if that
other equipment complies with CISPR EMISSION requirements.
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BP742N including AC-adapter should
assure that it is used in such environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The OMRON BP742N including AC-adapter uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11 Class B
The OMRON BP742N including AC-adapter is suitable for
use in all establishments, including domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BP742N including AC-adapter should
assure that it is used in such environment.
Immunity test IEC 60601 test
level Compliance level Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply inputlines
IEC 61000-4-11
<5 % UT (>95 % dip
in UT)
for 0.5 cycle
<5 % UT (>95 % dip
in UT)
for 0.5 cycle
Mains power quality should be
that of a typical commercial
and/or hospital environment. If the
user of the OMRON BP742N
including AC-adapter requires
continued operation during power
mains interruption, it is
recommended that the OMRON
BP742N including AC-adapter be
powered from an uninterruptible
power supply.
40 % UT (60 % dip
in UT)
for 5 cycles
40 % UT (60 % dip
in UT)
for 5 cycles
70 % UT (30 % dip
in UT)
for 25 cycles
70 % UT (30 % dip
in UT)
for 25 cycles
<5 % UT (95 % dip
in UT)
for 5 sec.
<5 % UT (95 % dip
in UT)
for 5 sec.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BP742N including AC-adapter should
assure that it is used in such environment.
Immunity test IEC 60601 test
level
Compli
ance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V rms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 V rms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the OMRON
BP742N including AC-adapter including cables, than the
recommended separation distance calculated from the
equation appropriate to the frequency of the transmitter.
Recommend separation distance
d = 1.2 √P
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as determined
by an electromagnetic site survey,
a should be less than
the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones
and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the OMRON BP742N including AC-adapter is used exceeds the applicable RF compliance
level above, the OMRON BP742N including AC-adapter should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the OMRON BP742N including AC-adapter.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distance between portable and mobile RF communications
equipment
and the OMRON BP742N including AC-adapter
OMRON BP742N including AC-adapter is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of this OMRON BP742N
including AC-adapter can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the OMRON
BP742N including AC-adapter as recommended below, according to the maximum output power of
the communications equipment.
Output Power of Transmitter in
Watt
Separation distance according to frequency of transmitter
in meter
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5GHz
d = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Manufactured for: OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
Distributed by: OMRON HEALTHCARE, INC.
1925 West Field Court Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2013 OMRON HEALTHCARE, INC.
Made in Vietnam
Especificaciones del producto
Marca: | Omron |
Categoría: | monitor de presión arterial |
Modelo: | BP742N |
¿Necesitas ayuda?
Si necesitas ayuda con Omron BP742N haz una pregunta a continuación y otros usuarios te responderán
monitor de presión arterial Omron Manuales
20 Octubre 2024
4 Octubre 2024
14 Septiembre 2024
13 Septiembre 2024
12 Septiembre 2024
12 Septiembre 2024
12 Septiembre 2024
3 Septiembre 2024
11 Agosto 2024
8 Agosto 2024
monitor de presión arterial Manuales
- monitor de presión arterial Nokia
- monitor de presión arterial Topcom
- monitor de presión arterial AnD
- monitor de presión arterial Aponorm
- monitor de presión arterial Thomson
- monitor de presión arterial Innoliving
- monitor de presión arterial Trisa
- monitor de presión arterial Pyle
- monitor de presión arterial OBH Nordica
- monitor de presión arterial Beurer
- monitor de presión arterial Melissa
- monitor de presión arterial Sencor
- monitor de presión arterial Lanaform
- monitor de presión arterial Medisana
- monitor de presión arterial Citizen
- monitor de presión arterial Cresta
- monitor de presión arterial Terraillon
- monitor de presión arterial Homedics
- monitor de presión arterial Sanitas
- monitor de presión arterial Rossmax
- monitor de presión arterial Microlife
- monitor de presión arterial Welch Allyn
- monitor de presión arterial Laica
- monitor de presión arterial Camry
- monitor de presión arterial Boso
- monitor de presión arterial Konig
- monitor de presión arterial Mx Onda
- monitor de presión arterial IHealth
- monitor de presión arterial Etekcity
- monitor de presión arterial Ecomed
- monitor de presión arterial Magic Care
- monitor de presión arterial Elta
- monitor de presión arterial Medel
- monitor de presión arterial Geratherm
- monitor de presión arterial Bioland
- monitor de presión arterial Orava
- monitor de presión arterial Levita
- monitor de presión arterial Scala
- monitor de presión arterial Silvergear
- monitor de presión arterial Uebe
- monitor de presión arterial Duronic
- monitor de presión arterial MTLogic
- monitor de presión arterial Hartmann
- monitor de presión arterial SunTech
- monitor de presión arterial Gima
- monitor de presión arterial Oro-Med
- monitor de presión arterial A&D
- monitor de presión arterial Dr. Senst
Últimos monitor de presión arterial Manuales
18 Octubre 2024
16 Octubre 2024
4 Octubre 2024
4 Octubre 2024
2 Octubre 2024
2 Octubre 2024
27 Septiembre 2024
27 Septiembre 2024
27 Septiembre 2024
27 Septiembre 2024