Differential Pressure Transmitter
PCE-DDMU 10
INSTRUCTION MANUAL
Table f ontents o c
1 General nformation. i
2 Product dentification. i
3 Installation.
4.Initial tart s -up
5 Operation.
6 Dimensions Fixing. /
7 Maintenance.
8 Return.
9 Warranty conditions.
Operating Manual
Dierential Pressure Transmitter PCE-DDMU 10
Menu button
1 General information
Do not blow into the pressure ports!
This causes damage to the device.
- This operating manual contains important informa-
tion on proper usage of the device. Read thisthe
operating manual carefully before installing and
starting up the pressure measuring device.
- c-Adhere to the safety notes and operating instru
tions which are given in the operating manual. Addi-
tionally applicable regulations regarding occupa-
tional safety, accident prevention as well as national
installation standards and engineering rules must
be complied with!
- This operating manual is part of the device. It must
be kept near its location and always be accessible
to all employees.
- This operating manual is copyrighted. The contents
of this operating manual reflect the version available
at the time of printing.
WARNING! To avoid operator hazards and dam-
age to he device, the following instructions t may
only be carried out by qualified technical perso n-
nel.
Limitation of liability: In -case of non
observance of the operating manual, inappropri-
ate use, modifications or damage, no liability is
assumed and warranty claims will be excluded.
Symbols used:
: Warning
!: Caution
: Note
– T –echnical modifications reserved
2 Product identification
2.1 Intended use
-
-
T differential pressurhe PCE-DDMU 10 e
transmitter has developed for t been he
measurement of pressur difference dry, non e of -
aggressiv gases pressuriz aire and ed and an bec
us for wide r of different HVACed a ange
applications.
Its robust desi means it c us i the gn an be ed n
laboratory under industrial conditions. and
Preferred e e n areas of us ar i heating, ventilation
and ng ean air conditioni systems; cl room and
medical technology, filter technology draft and
meteri checksng .
It s he operator's responsibility to check and verify i t
the suitability f he device or intended o t f the applica-
tion. f any bts emain, ease contact our les I dou r pl sa
department n order o ensure proper se. i t u
PCE Instruments is not liable for any incorrect
selections and heir t effects!
WARNING! – Danger by inappropriate use!
2.2 Identification
- The device can be identified by its manufacturing
label. It provides the most important data. T he
product can be clearly identified by the ordering
code.
Fig. 1: Manufacturing label
!The manufacturing label may not be removed
from the device!
2 .3 Package contents
Please verify that all listed parts are included in the
delivery undamaged and comply with your order :
1 Differential ressure Transmitter P PCE-DDMU 10
1 Operating manual
2.4 Transport and storage
Permissible temperature for storage: 10 °C to + 70 °C-
For transport make sure that both pressure inputs of
differential pressure sensors are open. The air transport
of absolute pressure sensors should only be done in a
pressure compensated cabin.
2.5 Declaration of onformity c / CE
The delivered evice all egal equirements. d fulfills l r
T applicablehe directives, harmonised standards d an
documents are listed in the EC eclaration of onformity d c .
The operational safety is additionally confirm by the ed
CE si on he manufacturing label. gn t
3 Installation
3.1 Mounting nd afety a s instructions
WARNING! This device may only be installed by
qualified technical personnel who have read and
understood the operating manual!
WARNING! Install the device only when depres-
surized and current free- !
!Handle this device with care, both in ed and packag
unpackaged condition!
!No modifications/changes should be made on the
device.
!Do not throw the device!
!Do not use any force when installing the device, to
prevent damage of the device and the plant!
!Avoid of sources interference (transformers,
transmitters, motors etc.) or sources of heat in the
surrounding area .
!Shocks or vibrations at the mounting connection
can cause distortions to the output signals.
3.2 Installation steps
- Carefully remove the pressure measuring device
from its package and dispose of the package prop-
erly.
- Mount the device in a suitable location using both
straps. The corresponding dimensions are provided
under "6".
- Connections should be vertical, i. e. the pressure
connections should point downwards. The sensors
are calibrated in the factory for these mounting con-
ditions. In addition, the formation of condensation in
the pressure tubing of the sensor is also reduced.
- Use the appropriate hoses for the connection.
Fig. 2: PCE-DDMU 10
3.3 installation Electrical
WARNING! This device may only be installed by
qualified technical personnel who have read and
understood the operating manual!
- Open the top cover and establish the electrical
connection of the device according to the follow-
ing description. Screw the top cover onto the box
again.
- Guide externally the power supply line through
the right so that the wires can cable fitting be
connected with the right terminal block (SUP-
PLY) without any problem.
- Next, tighten the cable gland by hand. Make
sure, however, that the cable is strain-relieved.
3.4 Pressure port
- Connect the hose with the higher pressure to the
right hand (positive) pressure port.
- Connect the hose with the lower pressure to the
left hand (negative) pressure port.
4 - Initial start up
After switching on the power supply, the output
signal can be measured. Variations in the output
signals may have two possible causes:
1. The warm up time of the sensor is about 30min- .
After this period, the sensor signal should be
stable for zero differential pressure and constant
ambient temperature.
2. For small pressure ranges, a slight deviation in
the zero point due to ambient conditions can oc-
cur. This error can be corrected by adjusting the
zero-point potentiometer of the sensor after the
warm- up time. (Set the output signal of the sen-
sor with both pressure inputs open to the nomi-
nal value.)
5 Operation
5 .1 General information
WARNING! - check Before start up, the user must
to ensure proper installation and for any visible de-
fects.
WARNING! The device may only be started and
operated by authorized personnel who have read
and understood the operating manual!
WARNING! The device may only be used within
the technical specifications the data in the (see
data sheet)!
WARNING! – Do not touch the bonding wires this
ca . n destroy the device
5.2 Setting of damping
P otentiometer for setting the damping is located
adjacent to the display on the left (see image 2).
The damping of the device can be set by turning a
size 2 Phillips screwdriver in the area of 0 to
5000ms.
5.3 Menu buttons
Zeroing: secondpush left menu button for at least 1
5.4 Configuration switch
3- - :wire system
1 2 3 4
Off / 0 ... 10 V 0 ... 20 mA
On 5 V0 ... / 4 ... 20 mA
Off Automatic zero adjustment
off
On Zero adjustment active at
start and for 24h
Nominal pressure range
Off Off 4 1,6 10 40 250 1000
Customized ranges
6 On Off 1,0 2,5 25 60 400
Off On 160 600
2- -wire system:
1 2 3 4
Off Off Automatic zero adjustment
off
On O ff Zero adjustment active at
start
O ff On Zero adjustment active at
start and for 24h
On On Zero adjustment active at
start, then every 7 days
Nominal pressure range
Off Off 4 1,6 10 40 250 1000
Customized ranges
On Off 6 1,0 2,5 25 60 400
Off On 160 600
The switches 1 and 2 don't have any function
at special pressure ranges.
terminal block
2- -wire system
Ordering code
Connector pinout
Type designation
Supply
Nominal
pressure
range
Sig nal
Serial
number
Display
(optional)
Potentiometer for adjust-
ing damping
P - hillips screwdriver
size 2
Configuration
switching
Negative
pressure connection
Positive
pressure connection
Cable gland M16x1.5
terminal block
3 wire system
3- -s 3- -wire ystem )(current / voltage wire system ( )current / voltage
w ith two outputs
supply terminal-: 7 - : VS
supply +: terminal 8: VS+
signal: 9 terminal : IOUT
signal: terminal 10: VOUT
Output 2 NO2: terminal 1 : NO2
Output 2 C2: terminal 2 : C2
Output 2 NC2 terminal 3 : NC2
Output 1 NO1 terminal 4 : NO1
Output 1 C1 terminal 5 : C1
Output 1 NC1 terminal 6 : NC1
supply -: terminal 7: VS-
supply +: terminal 8: VS+
signal: terminal 9 : IOUT
signal: terminal 10: VOUT
2- -s 2- - wire ystem (current) wire system (current) with two outputs
supply +: terminal 1: VS+
supply terminal-: 2 - : VS
supply +: terminal 1 + : VS
supply terminal-: 2 - : VS
output 1: terminal 3: S1
output 2: terminal 4: S2
Especificaciones del producto
Marca: | PCE Instruments |
Categoría: | Equipo de medición |
Modelo: | PCE-DDMU 10 |
¿Necesitas ayuda?
Si necesitas ayuda con PCE Instruments PCE-DDMU 10 haz una pregunta a continuación y otros usuarios te responderán
Equipo de medición PCE Instruments Manuales
24 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
Equipo de medición Manuales
- Equipo de medición Bosch
- Equipo de medición Panasonic
- Equipo de medición DeWalt
- Equipo de medición Honeywell
- Equipo de medición Pioneer
- Equipo de medición Alecto
- Equipo de medición Amprobe
- Equipo de medición Silverline
- Equipo de medición Gude
- Equipo de medición Black And Decker
- Equipo de medición Beurer
- Equipo de medición Yato
- Equipo de medición Megasat
- Equipo de medición Milwaukee
- Equipo de medición Extech
- Equipo de medición Victron Energy
- Equipo de medición ABB
- Equipo de medición Hazet
- Equipo de medición Metrix
- Equipo de medición Benning
- Equipo de medición Einhell
- Equipo de medición Voltcraft
- Equipo de medición Mitsubishi
- Equipo de medición Velleman
- Equipo de medición Vemer
- Equipo de medición Perel
- Equipo de medición Kyoritsu
- Equipo de medición IFM
- Equipo de medición Ideal
- Equipo de medición Flir
- Equipo de medición Grohe
- Equipo de medición Technoline
- Equipo de medición Soler And Palau
- Equipo de medición Continental Edison
- Equipo de medición Rossmax
- Equipo de medición Carrier
- Equipo de medición Testboy
- Equipo de medición Goclever
- Equipo de medición Danfoss
- Equipo de medición Hager
- Equipo de medición Bresser
- Equipo de medición TFA
- Equipo de medición Powerfix
- Equipo de medición Owon
- Equipo de medición Flex
- Equipo de medición Finder
- Equipo de medición Fluke
- Equipo de medición Fantini Cosmi
- Equipo de medición Eltako
- Equipo de medición Sekonic
- Equipo de medición EBERLE
- Equipo de medición Basetech
- Equipo de medición Be Cool
- Equipo de medición Laserliner
- Equipo de medición REV
- Equipo de medición Testo
- Equipo de medición Clean Air Optima
- Equipo de medición Eastron
- Equipo de medición ORNO
- Equipo de medición RIDGID
- Equipo de medición Multimetrix
- Equipo de medición GW Instek
- Equipo de medición Logilink
- Equipo de medición SIIG
- Equipo de medición Reely
- Equipo de medición Muller
- Equipo de medición Werma
- Equipo de medición Greisinger
- Equipo de medición Beha-Amprobe
- Equipo de medición Kopp
- Equipo de medición Kern
- Equipo de medición Kogan
- Equipo de medición Kemo
- Equipo de medición Peak
- Equipo de medición Uni-T
- Equipo de medición H-Tronic
- Equipo de medición Elgato
- Equipo de medición Hikmicro
- Equipo de medición PeakTech
- Equipo de medición Sonel
- Equipo de medición Tru Components
- Equipo de medición Emko
- Equipo de medición Entes
- Equipo de medición KS Tools
- Equipo de medición Trumeter
- Equipo de medición Delta Ohm
- Equipo de medición Atmel
- Equipo de medición Qualita
- Equipo de medición MGL Avionics
- Equipo de medición Adwa
- Equipo de medición Chauvin Arnoux
- Equipo de medición Megger
- Equipo de medición AREXX
- Equipo de medición Gossen Metrawatt
- Equipo de medición CEM
- Equipo de medición SHX
- Equipo de medición Hameg
- Equipo de medición TDE Instruments
- Equipo de medición Joy-it
- Equipo de medición Circutor
- Equipo de medición Wachendorff
- Equipo de medición Klein Tools
- Equipo de medición Duro Pro
- Equipo de medición Aerospace Logic
- Equipo de medición GMW
- Equipo de medición Janitza
- Equipo de medición Sanwa
- Equipo de medición MSW
- Equipo de medición Weidmüller
- Equipo de medición Qubino
- Equipo de medición HT Instruments
- Equipo de medición P3 International
- Equipo de medición Beckmann & Egle
- Equipo de medición Seek Thermal
- Equipo de medición INFICON
- Equipo de medición Motrona
- Equipo de medición Tempo
- Equipo de medición Camille Bauer
- Equipo de medición Blebox
- Equipo de medición Seaward
- Equipo de medición HOZO Design
- Equipo de medición Corentium
- Equipo de medición GQ
- Equipo de medición Tektronix
- Equipo de medición Dostmann Electronic
- Equipo de medición Murideo
- Equipo de medición SecuTech
- Equipo de medición NetPeppers
- Equipo de medición SensoProtect
- Equipo de medición Water-i.d.
- Equipo de medición ZKETECH
- Equipo de medición LabNation
- Equipo de medición Electronics International
- Equipo de medición Cliff
- Equipo de medición ClimeMET
- Equipo de medición Fixpoint
- Equipo de medición Kübler
- Equipo de medición AkYtec
- Equipo de medición Stahlwille
- Equipo de medición MASIMO
- Equipo de medición Milesight
- Equipo de medición Enda
- Equipo de medición CSL
- Equipo de medición Ermenrich
- Equipo de medición PICO
- Equipo de medición Siglent
- Equipo de medición Noyafa
- Equipo de medición Stamos
Últimos Equipo de medición Manuales
26 Octubre 2024
25 Octubre 2024
25 Octubre 2024
22 Octubre 2024
22 Octubre 2024
21 Octubre 2024
20 Octubre 2024
15 Octubre 2024
15 Octubre 2024
15 Octubre 2024