PCE Instruments PCE-DMM 60 Manual de Usario
PCE Instruments
Equipo de medición
PCE-DMM 60
Lee a continuación 📖 el manual en español para PCE Instruments PCE-DMM 60 (2 páginas) en la categoría Equipo de medición. Esta guía fue útil para 4 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios
Página 1/2
User Manual
Pressure Gauge PCE-DMM 70 / PCE-DMM 60
1.1 Information on the operating manual
This operating manual contains important information on proper
usage of the device. Read this operating manual carefully before
installing and starting up the pressure measuring device.
Adhere to the safety notes and operating instructions which are
given in the operating manual. Additionally applicable regulations
regarding occupational safety, accident prevention as well as
national installation standards and engineering rules must be
complied with!
This operating manual is part of the device and must be kept at
a, for the personnel always accessible location, close to the
installation position of the device.
This operating manual is copyrighted. The contents of this
operating manual reect the version available at the time of
printing. It has been issued to our best knowledge and belief.
However, errors may have occurred. For incorrect statements
and their consequences, liability cannot be assumed by PCE
Instruments.
– –Technical modications reserved
1.2 Symbols used
DANGER! Dangerous situation, which may result in –
serious or fatal injuries
WARNING! Potentially dangerous situation, which may –
result in fatal or serious injuries
CAUTION! Potentially dangerous situation, which may –
result in minor injuries
!CAUTION! Potentially dangerous situation, which may –
result in damage of objects
NOTE Tips and information for the user to ensure good–
conditions for the operation.
1.3 Target group
WARNING! To avoid hazards for the operator and damages
of the device, following described instructions have to be
worked out by qualied technical personnel.
1.4 Limitation of liability
No liability is assumed and warranty claims are excluded in case
of non-observance of the operating manual, improper application,
modication of or damage to the device.
1.5 Intended use
- The battery powered digital gauge has been designed for
extremely high demands in the sector of calibration and test
technology. can be easily and quickly installed in situ. It
- It is in the responsibility of the user to verify whether the
chosen device is suitable for the intended application. In case
of any doubts, contact our sales department to eliminate any
indistinctness. PCE Instruments does not assume any liability
for an incorrect selection and its consequences!
WARNING! - Danger through improper usage !
1.6 Package contents
Please verify that all listed parts are undamaged included in the
delivery and check for consistency specied in your order.
The batteries are already used. The circuit is interrupted by an
insulation foil. Take this before rst introduction, see in addition
battery change!!
2 Mechanical Installation.
2.1 Mounting and safety instructions
WARNING! Mount the device (pressure transmitter module)
always in the state without pressure and apart from the dis-
play!
WARNING! This device may only be installed by qualied
technical personnel who has read and understood the
operating manual!
WARNUNG! Do not use the display to tighten or solve to
the mechanical connection of the pressure transmitter mod-
ule!
!Handle this electronic precision measuring device carefully
in packed as well as in unpacked condition!
!The device must not be subject to any changes or
modications.
!The device may not be thrown!
!To avoid damaging the diaphragm, remove packaging and
protective cap only directly before starting up the device. A
delivered protective cap must be stored!
!Place the protective cap on the pressure port again
immediately after disassembling.
!Handle the unprotected diaphragm very carefully - it is very
sensitive and may be easily damaged.
!Do not use any force when installing the device to prevent
damage of the device and the plant!
Take note that no inadmissibly high mechanical stresses
occur at the pressure port as a result of the installation,
since this may cause a shifting of the characteristic curve or
to the demage. This is especially important for very small
pressure ranges as well as for devices with a pressure port
made of plastic.
In hydraulic systems, position the device in such a way that
the pressure port points upward (venting).
Provide a cooling line when using the device in steam lines.
2.2 General installation steps
- Carefully remove the pressure measuring device from the
package and dispose of the package properly.
- Then go ahead as detailed in the specic instructions
below.
2.3 Installation steps for DIN 3852
DO NOT USE ANY ADDITIONAL SEALING MATERIALS,
LIKE YARN, HEMP OR TEFLON TAPE!
- Check to ensure the proper groove fitting of the o-ring and
additionally to ensure no damage to the o-ring.
- Ensure that the sealing surface of the taking part is perfectly
smooth and clean. (R
Z 3.2)
- Screw the device into the corresponding thread by hand.
- If you have a device with a knurled ring, the transmitter has
to be screwed in by hand only.
- Devices with a spanner flat have to be fully tightened with
an open-end Devices (G1/4": approx. 5 Nm; G1/2": approx.
10 Nm ).
-The indicated tightening torques must not be
exceeded!
2.4 Installation steps for EN 837
- Use a suitable seal, corresponding to the medium and the
pressure input (e. g. a cooper gasket).
- Ensure that the sealing surface of the taking part is perfectly
smooth and clean. (R
Z 6.3)
- Screw the device into the corresponding thread by hand.
- Tighten it with a wrench (for G1/4": approx. 20 Nm; for
G1/2": approx. 50 Nm).
-The indicated tightening torques must not be
exceeded!
2.5 Installation steps for NPT
- Use a suitable seal (e. g. a PTFE-strip).
- Screw the device into the corresponding thread by hand.
- Tighten it with a wrench (for 1/4" NPT: approx 30 Nm; for.
1/2" NPT: approx. 70 Nm).
-The indicated tightening torques must not be
exceeded!
2 Installation steps for internal threads M20x1.5 and 9/16" .6
UNF (for DM01-500 HD)
- Screw the high pressure connection into the internal thread
of the pressure port and tighten it properly with approx. 160
Nm.
-DANGER! The high pressure tube seals metal- -metal to
in the chamfer of the pressure port. No further seal is
allowed with this high pressure connection. A wrong
installation can cause enormous danger!
3 Connection display with pressure transmitter module.
Connect display with module as follows: pressure transmitter
- bring together carefully the display with pressure transmit-
ter module.
- press the display sturdy module to this
pressure transmitter
engages.
4 . Supply / changing the batteries
As soon as in the display the announcement of "battery" is
shown, carry out battery change as follows:
-unscrew three xing screws with a suitable screwdriver.
-take the battery case cap and exchange the batteries 3 x
1.5 V AA (remove the insulation foil before first intro-
duction).
-lock the device after that properly
!An incorrect usage may cause a leak out of batteries and
so a damage the !
device
!Never combine batteries of dierent types or old with new
ones!
!Make sure that the batteries are connected correctly with
the corresponding contacts in the battery tray.
!Never try to charge batteri es, demount them, or short -
circuit them.
!use only batteries with UL certication
!Keep the batteries away from heat and unshielded ame.
5. Data logger
The battery powered digital gauge disposes of an integrated data
logger. The measuring values stored away in the device can be
selected above the communication interface by means of
software.
5.1 PC - connection
Connect with a computer as follows: device
-unscrew the protective screwing of the communication
interface with a suitable slit screwdriver.
-connect the handle plug of the connecting cable (included
in delivery) with the interface socket of the . Codevice n-
nect the side with the USB plug with a free USB connec-
tion on the computer.
-install the COM dri ver and data logger software .
-after the use, disconnect the connection and lock the
protection screwing again properly.
6 Initial start- . up
remove the insulation foil before rst introduction .
WARNING! Before start-up, the user has to check for
proper installation and for any visible defects.
WARNING! The device can be started and operated by
authorized personnel only, who have read and understood
the operating manual!!
WARNING! The device has to be used within the technical
specications, only (compare the data in the data sheet)!
7 Placing out of service .
WARNING! When dismantling the device, it must always be
done in the depressurized and currentless condition! Check
also if the medium has to be drained o before dismantling!
WARNING! Depending on the medium, it may cause
danger for the user. Comply therefore with adequate
precautions for purication.
8 Operation .
8.1 Operating and display elements
The display of the measuring value as well as configuring the
single parameters occurs menu-steered about a LC display
capable of graphic arts. The single functions are regulated on the
basis of three-front-sided arranged push buttons.
The menu system is closed, thereby one can "browse" forward
as well as backward by the single setting menus to reach to the
desired setting point.
8.2 Structure of the menu system
Fig 3 Display and operating foil .
b ograph ar
display
right button
menu button
left button
Fig. 2 Battery case cap and communication interface
battery case
battery case cap
xing screw
connecting cable
comm. -
interface
insulation foil
protection screwing
battery
timer / battery
pressure value
time
temperature
Fig 1 . Lock against protection
lock against
protection
pressure transmit-
ter module
display
PCE Americas Inc.
711 Commerce Way
Suite 8
Jupiter
FL-33458
USA
From outside US: +1
Tel: (561) 320-9162
Fax: (561) 320-9176
info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Unit 11
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@pce-instruments.co.uk
- if is a functional button and can be congured in the menu 6. Light-, Zero- - or Min- func-Left uttonB : , Reset
tion can be assigned to the button. The congured function is active in the display mode. Hold
the button during approx. 2 seconds to activate the function. In the operating mode you move in
the menu system backward or you reduce the set value. "<<"
- if is a functional button and can be congured in the menu 6 Light-, Zero- - or Max- func-Right Button: . , Reset
tion can be assigned to the button . Hold the button during approx. 2 seconds to activate the .
function. In the operating mode you move in the menu system forward ">>" or you raise the set
value.
- : f this button "menu" you reach in the operating mode; moreover, she serves for Menu-Button by pressure o
the choice of the single menu points "Edit" or in the conrmation the op posed worth "Next".
To the conguration of the single menu points the desired menu point is to be put with the help of the left button "<<"
or right button ">>". Conrm this afterwards with the menu button "Edit" the menu point is marked and the congura-
tion can begin.
To store an opposed value the menu button "Next" must be also pressed. To leave the menu press the menu button
so often to the mark of the single menu point and choose with the left button or right button the menu 1 "<<" ">>"
and press the menu button once more "Exit". The operating mode will also leave after approx. 1 min automatically.
Changes become rst after activity of the menu button "Next" and after abandonment of the menu point eec-
tive. With the abandonment of the whole menu system the opposed parameters are checked once again in d
e-
pendence to each other and concerning the characteristics of the device. With the conguration of the unity a
conversion of the measuring area occurs in the new unity only after abandonment of the menu system. Accord-
ing to pressure area not all units can be also used if necessary.
If the device came in contact with hazardous
substances, certain precautions have to be
complied with for purication!
9. Disposal
The device must be disposed according to the
European Directives 2002/96/EG and 2003/108/EG
(on waste electrical and electronic equipment)
Waste of electrical and electronic equipment may
not be disposed by domestic refuse!
WARNING! Depending on the measuring medium, deposit
on the device may cause danger for the user and the
environment. Comply with adequate precautions for
purication and dispose of it properly.
10. Warranty conditions
The warranty conditions are subject to the legal warranty period
of 24 months from the date of delivery. In case of improper use,
modications of or damages to the device, we do not accept
warranty claims. Damaged diaphragms will also not be accepted.
Furthermore, defects due to normal wear are not subject to
warranty services
11. Declaration of conformity / CE
The delivered device fulls all legal requirements. The applied
directives, harmonised standards and documents are listed in the
EC declaration of conformity
Additionally, the operational safety is conrmed by the CE sign
on the manufacturing label.
warranty services.
12. Maintenance
In principle, this device is maintenance-free. If desired, the
housing of the device can be cleaned when switched of using a
damp cloth and non-aggressive cleaning solutions.
With certain media, however, the diaphragm may be polluted or
coated with deposit. It is recommended to dene corresponding
service intervals for control. After placing the device out of
service correctly, the diaphragm can usually be cleaned carefully
with a non-aggressive cleaning solution and a soft brush or
sponge. If the diaphragm is calcied, it is recommended to send
the device to PCE Instruments for decalcication. Please read
therefore the chapter “Repair” below.
!A false cleaning of the device can cause an irreparable
damage on the diaphragm. Therefore never use pointed ob-
jects or pressured air for cleaning the diaphragm.
13. Service / Repair
We recommend that the instrument is regularly recalibrated by
PCE Instruments, with time intervals of approx. 12 months.
Before you return, the device has to be cleaned carefully and
packed shatter-proofed. You have to enclose a notice of return
with detailed defect description when sending the device. Contact
PCE Instruments to ask for this form. If your device came in
contact with harmful substances, a declaration of
decontamination is additionally required. Should you dispatch a
device without a declaration of decontamination and if there are
any doubts in our service department regarding the used
medium, repair will not be started until an acceptable declaration
is sent.
14. Menu list
Switch on
The device can be switched on with every button (left button, menu button, right button).
Menu 1/9
on / off
By push the menu button the operating mode will be selected / left / ("Menu" / "Exit").
Menu 2/9
date / time / info
Setting of topical time and the topical date. Software version will be shown.
Menu 3/9
Units
Setting of the pressure unit
adjustable units: [bar], [mbar], [PSI], [inHg], [cmHg], [mmHg], [hPa], [kPa], [MPa], [mmH2O], [mH2O], [kg/cm
2]
or [user] (the user-dened unit can be programmed only by means of the software BD|LOG), a conversation of
all pressure related parameters is carried out automatically
Setting of decimal places
adjustable decimal place: standard [Std], [+1], [+2]
Setting of the temperature unit
the unity is not adjustable. Factory setting [°C]
Menu 4/9
Min / Max Values
Language: Setting of user languages German [DE] or English [EN]
Measuring rate: here it is put how often the measurement is carried out and is indicated.
Possible settings are: a measurement per second [1/sec]
or two measurements per second [2/sec]
Auto O Time: Setting of the automatic switch o in minutes. The automatic switch o is able in steps from
[1 min], [2 min], [3 min], [4 min] or [5 min]; 30 seconds before switching o the timer is
activated and indicated in the display.
By the option [O] the device can be deactivated. After deactivation, the precision digital
gauge is in the continuous operation.
Menu 5/9
Configuration 1
Language: Setting of user languages German [DE] or English [EN]
Measuring rate: here it is put how often the measurement is carried out and is indicated.
Possible settings are: a measurement per second [1/sec]
or two measurements per second [2/sec]
Auto O Time: Setting of the automatic switch o in minutes. The automatic switch o is able in steps from
[1 min], [2 min], [3 min], [4 min] or [5 min]; 30 seconds before switching o the timer is
activated and indicated in the display.
By the option [O] the device can be deactivated. After deactivation, the precision digital
gauge is in the continuous operation.
Menu 6/9
Configuration 2
Button settings: left button / right button
Left Button: function congurations: [Min], [Light], [Zero], [Reset]
Right Button: function congurations: [Max], [Light], [Zero], [Reset]
Function descriptions:
- [Min] / [Max] minimum / maximum pressure value will be showed
- [Light] the display backlight is activated
- [Zero] the zero point is automatically put, in the display appears
- [Reset] the opposed zero is put back, goes out
Damping: the damping can be put in 1 second steps between [1 sec] and [10 sec] or be
deactivated by option [O]
Menu 7/9
Backlight
Backlight settings of display
On Time: the backlight duration can be put in 1 second steps between [1 sec] and [10 sec] or be
deactivated by option [O]
Brightness: the brightness can be put into 5% of steps between [0%] and [100%]
Menu 8/9
Data Logger
Data Logger conguration
Data Logger: the following settings are possible: linearly [Linear] (value admission to the counter level 8500
is reached), cyclically ( [Loop] (after the value is reached in 8500, the data logger automatically
begins the values once more to grasp and, besides, headlines the old values or [O] )
(in the display appears if the data logger is activated and goes out if the data logger is o)
Interval: Intervals to the memory of the measuring values (pressure / temperature) second [1- sec],99
minute [1- min], hour [1-99 hour] or day [1-99 ys] 99 da
Time: Measuring value admission: in which time the measuring value admission should occur (is
only eciently for the setting "daily" ).
Counter: Number of the grasped measuring values is indicated. Maximum value: 8500
(Put back counter reading: menu point with "Edit" select → button "Next" four times operate → button ">>" operate. There appears the
question "Reset?" → once more operate the button ">>". It seems "Sure?" additional confirmation whether the value should be reset →
repeated conrming with the button ">>" reset the grasped measuring values Display announcement "Counter: 0/8500").
Menu 9/9
Sensor Data
[SN:] the standard number is indicated [Upper] Measuring area end
[Lower] Measuring area beginning [Date] Production date
Values are put by factory and are not adjustable.
Error
Display "No Sensor": display and pressure sensor module are separated.
Especificaciones del producto
Marca: | PCE Instruments |
Categoría: | Equipo de medición |
Modelo: | PCE-DMM 60 |
¿Necesitas ayuda?
Si necesitas ayuda con PCE Instruments PCE-DMM 60 haz una pregunta a continuación y otros usuarios te responderán
Equipo de medición PCE Instruments Manuales
24 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
Equipo de medición Manuales
- Equipo de medición Bosch
- Equipo de medición Panasonic
- Equipo de medición DeWalt
- Equipo de medición Honeywell
- Equipo de medición Pioneer
- Equipo de medición Alecto
- Equipo de medición Amprobe
- Equipo de medición Silverline
- Equipo de medición Gude
- Equipo de medición Black And Decker
- Equipo de medición Beurer
- Equipo de medición Yato
- Equipo de medición Megasat
- Equipo de medición Milwaukee
- Equipo de medición Extech
- Equipo de medición Victron Energy
- Equipo de medición ABB
- Equipo de medición Hazet
- Equipo de medición Metrix
- Equipo de medición Benning
- Equipo de medición Einhell
- Equipo de medición Voltcraft
- Equipo de medición Mitsubishi
- Equipo de medición Velleman
- Equipo de medición Vemer
- Equipo de medición Perel
- Equipo de medición Kyoritsu
- Equipo de medición IFM
- Equipo de medición Ideal
- Equipo de medición Flir
- Equipo de medición Grohe
- Equipo de medición Technoline
- Equipo de medición Soler And Palau
- Equipo de medición Continental Edison
- Equipo de medición Rossmax
- Equipo de medición Carrier
- Equipo de medición Testboy
- Equipo de medición Goclever
- Equipo de medición Danfoss
- Equipo de medición Hager
- Equipo de medición Bresser
- Equipo de medición TFA
- Equipo de medición Powerfix
- Equipo de medición Owon
- Equipo de medición Flex
- Equipo de medición Finder
- Equipo de medición Fluke
- Equipo de medición Fantini Cosmi
- Equipo de medición Eltako
- Equipo de medición Sekonic
- Equipo de medición EBERLE
- Equipo de medición Basetech
- Equipo de medición Be Cool
- Equipo de medición Laserliner
- Equipo de medición REV
- Equipo de medición Testo
- Equipo de medición Clean Air Optima
- Equipo de medición Eastron
- Equipo de medición ORNO
- Equipo de medición RIDGID
- Equipo de medición Multimetrix
- Equipo de medición GW Instek
- Equipo de medición Logilink
- Equipo de medición SIIG
- Equipo de medición Reely
- Equipo de medición Muller
- Equipo de medición Werma
- Equipo de medición Greisinger
- Equipo de medición Beha-Amprobe
- Equipo de medición Kopp
- Equipo de medición Kern
- Equipo de medición Kogan
- Equipo de medición Kemo
- Equipo de medición Peak
- Equipo de medición Uni-T
- Equipo de medición H-Tronic
- Equipo de medición Elgato
- Equipo de medición Hikmicro
- Equipo de medición PeakTech
- Equipo de medición Sonel
- Equipo de medición Tru Components
- Equipo de medición Emko
- Equipo de medición Entes
- Equipo de medición KS Tools
- Equipo de medición Trumeter
- Equipo de medición Delta Ohm
- Equipo de medición Atmel
- Equipo de medición Qualita
- Equipo de medición MGL Avionics
- Equipo de medición Adwa
- Equipo de medición Chauvin Arnoux
- Equipo de medición Megger
- Equipo de medición AREXX
- Equipo de medición Gossen Metrawatt
- Equipo de medición CEM
- Equipo de medición SHX
- Equipo de medición Hameg
- Equipo de medición TDE Instruments
- Equipo de medición Joy-it
- Equipo de medición Circutor
- Equipo de medición Wachendorff
- Equipo de medición Klein Tools
- Equipo de medición Duro Pro
- Equipo de medición Aerospace Logic
- Equipo de medición GMW
- Equipo de medición Janitza
- Equipo de medición Sanwa
- Equipo de medición MSW
- Equipo de medición Weidmüller
- Equipo de medición Qubino
- Equipo de medición HT Instruments
- Equipo de medición P3 International
- Equipo de medición Beckmann & Egle
- Equipo de medición Seek Thermal
- Equipo de medición INFICON
- Equipo de medición Motrona
- Equipo de medición Tempo
- Equipo de medición Camille Bauer
- Equipo de medición Blebox
- Equipo de medición Seaward
- Equipo de medición HOZO Design
- Equipo de medición Corentium
- Equipo de medición GQ
- Equipo de medición Tektronix
- Equipo de medición Dostmann Electronic
- Equipo de medición Murideo
- Equipo de medición SecuTech
- Equipo de medición NetPeppers
- Equipo de medición SensoProtect
- Equipo de medición Water-i.d.
- Equipo de medición ZKETECH
- Equipo de medición LabNation
- Equipo de medición Electronics International
- Equipo de medición Cliff
- Equipo de medición ClimeMET
- Equipo de medición Fixpoint
- Equipo de medición Kübler
- Equipo de medición AkYtec
- Equipo de medición Stahlwille
- Equipo de medición MASIMO
- Equipo de medición Milesight
- Equipo de medición Enda
- Equipo de medición CSL
- Equipo de medición Ermenrich
- Equipo de medición PICO
- Equipo de medición Siglent
- Equipo de medición Noyafa
- Equipo de medición Stamos
Últimos Equipo de medición Manuales
26 Octubre 2024
25 Octubre 2024
25 Octubre 2024
22 Octubre 2024
22 Octubre 2024
21 Octubre 2024
20 Octubre 2024
15 Octubre 2024
15 Octubre 2024
15 Octubre 2024