Kern HMO Manual de Usario

Lee a continuación 📖 el manual en español para Kern HMO (17 páginas) en la categoría Equipo de medición. Esta guía fue útil para 9 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios

Página 1/17
Sauter GmbH
Ziegelei 1
D-72336 Balingen
e-mail: info@kern-sohn.com
Phone : +49-[0]7433- 9933-0
Fax: +49-[0]7433- - 9933 149
Internet: www.sauter.eu
Instruction manual mobile Leeb
hardness tester
SAUTER HMO
Version 2.0
04/2020
GB
HMO- -e- BA 2020
PROFESSIONAL MEASURING
2 HMO- -e- BA 2020
GB
Thank you for purchasing the mobile digital Leeb impact hardness tester from
SAUTER. We hope that you will be very satised with the high quality of this device
and its extensive functionality. For any questions, wishes and suggestions we are at
your disposal.
Table of contents:
1 Before commissioning .................................................................................. 3
2 Introduction .................................................................................................... 3
2.1 Measuring principle .................................................................................................................. 3
2.2 Structure .................................................................................................................................... 4
2.3 Technical data ........................................................................................................................... 5
2.4 Area of application ................................................................................................................... 5
3 Test procedure ............................................................................................... 6
3.1 Preparation of the device 6........................................................................................................
3.2 Preparation of the sample ....................................................................................................... 6
3.3 Test execution .......................................................................................................................... 8
3.4 Browsing ................................................................................................................................... 8
3.5 Print 9............................................................................................................................................
4 Setup ............................................................................................................... 9
4.1 Settings in the main mask ....................................................................................................... 9
4.2 Menu settings ......................................................................................................................... 10
5 Care and maintenance ................................................................................. 14
5.1 Periodic maintenance ............................................................................................................ 14
5.2 Store ......................................................................................................................................... 14
5.3 Calibration ............................................................................................................................... 15
6 Annex ............................................................................................................ 16
6.1 Conversion area ..................................................................................................................... 17
6.2 Scope of delivery .................................................................................................................... 17
HMO- -e- 3 BA 2020
1 Before commissioning
Before putting the device into operation, check the delivery for any transport damage
to the packaging, the plastic case and the device itself. Should this be the case,
SAUTER must be contacted immediately.
2 Introduction
2.1 Measuring principle
The measuring principle for the measurements carried out with this dynamic impact
hardness tester is physically quite uncomplicated. An impact body with a hard metal
tip is hurled against the surface of the test object by the spring force. When the impact
body hits the surface, it can be damaged, which ultimately leads to a loss of kinetic
energy.
This energy loss is calculated on the basis of velocity measurements, at a certain point
in time when the impactor is at a certain distance from the surface, both in the rebound
and impact phases.
The xed magnet in the impact body generates an induction voltage in the simple wire
coil of the impact tester. The voltage of the signal is proportional to the speed of the
impact body. The signal is processed by the electronics so that the hardness value can
be read on the display and then stored.
English
German
Impact phase
Impact phase
Rebound phase
Rebound phase
Time
Time

Especificaciones del producto

Marca: Kern
Categoría: Equipo de medición
Modelo: HMO

¿Necesitas ayuda?

Si necesitas ayuda con Kern HMO haz una pregunta a continuación y otros usuarios te responderán




Equipo de medición Kern Manuales

Kern

Kern HMM Manual de Usario

19 Septiembre 2024
Kern

Kern HDA 100-1 Manual de Usario

19 Septiembre 2024
Kern

Kern HMO Manual de Usario

19 Septiembre 2024
Kern

Kern HO 2K Manual de Usario

19 Septiembre 2024
Kern

Kern HK-D Manual de Usario

19 Septiembre 2024
Kern

Kern HO 5K Manual de Usario

19 Septiembre 2024
Kern

Kern FS 4-500 Manual de Usario

19 Septiembre 2024
Kern

Kern FH 100KUK Manual de Usario

19 Septiembre 2024
Kern

Kern HK-DB Manual de Usario

19 Septiembre 2024

Equipo de medición Manuales

Últimos Equipo de medición Manuales