Safety Symbols
#Refer to the instructions in the manual.
Indicates instrument with double or reinforced insulation.
Indicates that this instrument can clamp on live bare
conductors when the voltage to be tested is below
Circuit - Ground-to-Earth voltage against the indicated
Measurement Category.
AC
DC
This instrument satisfies the marking requirement
defined in the WEEE Directive. This symbol indicates
separate collection for electrical and electronic
equipment.
○ Measurement Category
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environ-
ments, categorized as O to CAT.IV, and called measurement
categories. Higher-numbered categories correspond to electrical
environments with greater momentary energy, so a measuring
instrument designed for CAT.III environments can endure greater
momentary energy than one designed for CAT.II.
O : Circuits which are not directly connected to the mains
power supply.
CAT.II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT.III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT.IV : The circuit from the service drop to the service
en tr anc e, an d to th e po wer m ete r an d p rim ary
overcurrent protection device (distribution panel).
# DANGER
● Never make measurement on a circuit in which the electrical
potential exceeds 300V in order to avoid possible shock
hazard.
● Do not make measurement when thunder is rumbling. If the
instrument is in use, stop the measurement immediately and
remove the instrument from the measured object.
● Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the instrument may
cause sparking, which can lead to an explosion.
● Put insulated protective gears when there is a danger of
electrical shock hazard.
● The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible
shorting where the measured object has exposed metal parts.
● Never attempt to use the instrument if it's surface or your
hand is wet.
Otherwise, electrical shock accident may occur.
● Do not exceed the maximum allowable input of any measuring
range.
● Never open the battery compartment cover and the instrument
case when making a measurement.
● The instrument is to be used only in its intended applications
or conditions. Otherwise, safety functions equipped with the
instrument doesn’t work, and instrument damage or serious
personal injury may be caused.
# WARNING
● Never attempt to make any measurement, if any abnormal
conditions are noted, such as broken case, and exposed
metal parts.
● Do not install substitute parts or make any modification to
the instrument. Return the instrument to your local Kyoritsu
distributor for repair or re-calibration in case of suspected
faulty operation.
● Always keep your fingers and hands behind the barrier on the
instrument to avoid the possible shock hazard.
● Do not try to replace batteries if the surface of the instrument
is wet.
Power off the instrument before opening the Battery
Compartment Cover for a battery replacement.
1. Safety Warnings
This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for Electronic
Measuring apparatus, and delivered in the best condition after
passing quality control tests.
This instruction manual contains warnings and safety rules which
have to be observed by the user to ensure safe operation of the
instrument and to maintain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
# WARNING
● Read through and understand instructions contained in this
manual before using the instrument.
● Keep the manual at hand to enable quick reference whenever
necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained in
the manual.
It is essential that the above instructions are adhered to.
Failure to follow the above instructions may cause injury and or
instrument damage.
The symbol indicated on the instrument, means that the user #
must refer to the related parts in the manual for safe operation of
the instrument. It is essential to read the instructions wherever
the symbol appears in the manual.
# DANGER is reserved for conditions and actions that are
likely to cause serious or fatal injury.
# WARNING is reserved for conditions and actions that can
cause serious or fatal Injury.
# CAUTION is reserved for conditions and actions that can
cause minor injury or instrument damage.
# CAUTION
● Do not step on or pinch the cord to prevent the jacket of cord
from being damaged.
● The output connector shall be removed or connected without
clamping a conductor. Otherwise, it may cause a failure.
● Do not expose the instrument to direct sunlight, high
temperatures, humidity or dew.
● Do not use this instrument in dusty place and where the
instrument is likely to get wet.
● Power off the instrument after use. Remove the batteries if
the instrument is to be stored and will not be in use for a long
period.
● Never give shocks, such as vibration or drop, which may
damage the instrument.
● Use a damp cloth and detergent for cleaning the instrument.
Do not use abrasives or solvents.
● Take sufficient care to avoid shock, vibration or excessive
force when handling the instrument. Otherwise, precisely
adjusted transformer jaws will be damaged.
● When transformer jaws do not fully close, never try to close
them by force, but make them free to move and try again. If a
foreign substance is stuck in the jaw tips, remove it.
● Do not open the jaws by force when transformer jaws are
frozen.
● Hold the inserting part (except for the cord) and disconnect
the Output Connector from the measuring instrument so as
not to cause a break in the cable.
2.Features
● Clamp sensor for AC/DC current measurements
● Designed to meet the international safety standard
IEC61010-2-032 CAT.III 300V, Pollution degree 2
3.Instrument layout
<Power on/ off>
One press of the Power Switch powers on the instrument and the
green LED lights up. Another press of the Power Switch powers
off the instrument. The instrument is automatically powered off
in about 20 min after the last switch operation. (Auto-power-off
Function)
<To disable the Auto-power-off Function>
Keep the Power Switch pressed down at least 3 sec and power
on the instrument. Then the red LED lights up and the Auto-
power-off Function doesn’t activate when 20 min pass after the
last switch operation.
4.Specification
Model KEW8115
Measuring Range AC0.1 130Arms〜
(185Apeak) DC0 180A〜
Output voltage AC/DC 10mV/A
Accuracy
(Input: sine wave)
(*1)
50/60Hz:
±(1.2%rdg+0.4mV)
40 1kHz:〜
±(2.5%rdg+0.4mV)
±(1.2%rdg+0.4mV)
(*2)
Temperature & humidity range
(Guaranteed accuracy)
23±5ºC, relative humidity 85% or less
(no condensation)
Operating temperature
& humidity range
-10 55 ºC , relative humidity 85% or less 〜
(no condensation)
Storage temperature &
humidity range
-30 70 ºC , relative humidity 85% or less 〜
(no condensation) (*3)
Temperature coefficient
(Accuracy at 23±5℃×±0.1%rdg)/℃ should be added.
(Temperature ranges: -10 to 18℃ and 28 to 55℃)
Output impedance Approx 10Ω or less
Location for use Altitude 2000m or less, indoor use
Applicable standards
IEC 61010-1 CAT. III 300V, Pollution degree 2
IEC 61010-2-032
IEC 61326-1(EMC), EN50581(RoHS)
Withstand voltage
AC3470V (RMS 50/60Hz) for 5 sec
between Jaw and enclosure
between enclosure and output terminal
between Jaw and output terminal
Insulation resistance
10MΩ or more/1000V
between Jaw and enclosure
between enclosure and output terminal
between Jaw and output terminal
Power source DC3V (size AAA alkaline battery LR03×2pcs)
Auto-power-off
Approx 20 min after powering on the instrument
Low battery warning 2.2±0.2V or less(*4)
red LED flash
Current consumption 25mA or less (battery voltage 3V)
Continuous use Approx 40 hours
Conductor size Max φ12mm
Cord length Approx 120cm
Dimension 127(L)×42(W)×22(D)mm
Weight Approx 140g
Accessories Instruction manual, battery, Soft case
(*1) Under the EMC environment, the instrument temporally
reads errors due to the influences of electrical magnetic
fields (about 15mV at 3V/m). Declared accuracy is
guaranteed when a conductor under test is at the center of
the Transformer jaws.
(*2) This instrument doesn’t have Zero Adjustment Function.
This accuracy is defined after a zero-adjustment by using
the device connected with this instrument. It is advisable to
do Zero adjustment before making measurements because
zero value may vary due to temperature change.
(*3) without batteries
(*4) powers off automatically at 1.9±0.2V or less
Instrument works properly even while Low battery indicator
is flashing.
5.Measurement
(1) Press the Power Switch to power on the instrument. The
green LED lights up. (*5)
(2) Connect the output terminal to the input terminal of the
measuring instrument.
(3) Press the trigger to open the transformer jaws, and clamp
onto a conductor.
(4)
Confirm that the tips of the transformer jaws are surely closed.
(5) Start a measurement. The measured conductor should be at
the center of the jaws. (*6)
(6) The measured result is displayed on the connected
measuring instrument.
(*5) Keep the Power Switch pressed down at least 3 sec and
power on the instrument to disable the Auto-power-off
Function. In this case, the red LED lights up after powering
on the instrument.
(*6) Triangle marks on the Transformer Jows showing the center
of the Jaws.
* Currents flowing from the top side (side with the Power
switch) to the bottom side (Battery Compartment Cover) are
positive, the ones flowing from the bottom side to the top side
is negative.
* Max conductor size for KEW8115 is max φ12mm. Accurate
measurements cannot be taken on conductors thicker than
above size since the transformer jaws don’t fully close.
*
Do not give big shocks, when opening / closing tranformer jaws.
6.Battery Replacement
Replace the batteries when a low batter indicator (red LED)
flashes on the LCD.
(1) Press the Power Switch and power off the instrument.
(2) Loosen 1 screw at the bottom side of the instrument to
remove the Battery Compartment Cover.
(3) Replace the batteries with new ones.
(size AAA alkaline battery : LR03×2 pcs)
(4) Install the Battery Compartment Cover and tighten the
screws.
# WARNING
● Do not try to replace batteries when making a measurement
to avoid electric shock hazard.
# CAUTION
● Do not mix batteries of different types or new batteries with
used ones.
● Make sure to install batteries in correct polarity as indicated in
the battery compartment.
Instruction Manual
AC/DC Clamp Sensor
KEW 8115
Incoming wire
O:Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring
LED
Transformer Jaw
Barrier
Trigger
Output Terminal
Power switchi
Cord
Green Light : power on
Red Light : no Auto-power-off Function
Red Flash : Low battery warning
φ4mm
Banana Plug
(
)
7-16 92-1969D
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.
Battery Compartment Cover
Screw
Conductor
Current direction
arrow
Especificaciones del producto
Marca: | Kyoritsu |
Categoría: | Equipo de medición |
Modelo: | KEW 8115 |
¿Necesitas ayuda?
Si necesitas ayuda con Kyoritsu KEW 8115 haz una pregunta a continuación y otros usuarios te responderán
Equipo de medición Kyoritsu Manuales
15 Septiembre 2024
11 Septiembre 2024
11 Septiembre 2024
11 Septiembre 2024
11 Septiembre 2024
11 Septiembre 2024
11 Septiembre 2024
10 Septiembre 2024
10 Septiembre 2024
10 Septiembre 2024
Equipo de medición Manuales
- Equipo de medición Bosch
- Equipo de medición Panasonic
- Equipo de medición DeWalt
- Equipo de medición Honeywell
- Equipo de medición Pioneer
- Equipo de medición Alecto
- Equipo de medición Amprobe
- Equipo de medición Silverline
- Equipo de medición Gude
- Equipo de medición Black And Decker
- Equipo de medición Beurer
- Equipo de medición Yato
- Equipo de medición Megasat
- Equipo de medición Milwaukee
- Equipo de medición Extech
- Equipo de medición Victron Energy
- Equipo de medición ABB
- Equipo de medición Hazet
- Equipo de medición Metrix
- Equipo de medición Benning
- Equipo de medición Einhell
- Equipo de medición Voltcraft
- Equipo de medición Mitsubishi
- Equipo de medición Velleman
- Equipo de medición Vemer
- Equipo de medición Perel
- Equipo de medición IFM
- Equipo de medición Ideal
- Equipo de medición Flir
- Equipo de medición Grohe
- Equipo de medición Technoline
- Equipo de medición Soler And Palau
- Equipo de medición Continental Edison
- Equipo de medición Rossmax
- Equipo de medición Carrier
- Equipo de medición Testboy
- Equipo de medición Goclever
- Equipo de medición Danfoss
- Equipo de medición Hager
- Equipo de medición Bresser
- Equipo de medición TFA
- Equipo de medición Powerfix
- Equipo de medición Owon
- Equipo de medición Flex
- Equipo de medición Finder
- Equipo de medición Fluke
- Equipo de medición Fantini Cosmi
- Equipo de medición Eltako
- Equipo de medición Sekonic
- Equipo de medición EBERLE
- Equipo de medición Basetech
- Equipo de medición Be Cool
- Equipo de medición Laserliner
- Equipo de medición REV
- Equipo de medición Testo
- Equipo de medición Clean Air Optima
- Equipo de medición Eastron
- Equipo de medición ORNO
- Equipo de medición RIDGID
- Equipo de medición Multimetrix
- Equipo de medición GW Instek
- Equipo de medición Logilink
- Equipo de medición SIIG
- Equipo de medición Reely
- Equipo de medición Muller
- Equipo de medición Werma
- Equipo de medición Greisinger
- Equipo de medición Beha-Amprobe
- Equipo de medición Kopp
- Equipo de medición Kern
- Equipo de medición Kogan
- Equipo de medición Kemo
- Equipo de medición Peak
- Equipo de medición Uni-T
- Equipo de medición H-Tronic
- Equipo de medición Elgato
- Equipo de medición Hikmicro
- Equipo de medición PeakTech
- Equipo de medición Sonel
- Equipo de medición Tru Components
- Equipo de medición Emko
- Equipo de medición Entes
- Equipo de medición KS Tools
- Equipo de medición Trumeter
- Equipo de medición Delta Ohm
- Equipo de medición Atmel
- Equipo de medición Qualita
- Equipo de medición MGL Avionics
- Equipo de medición Adwa
- Equipo de medición Chauvin Arnoux
- Equipo de medición Megger
- Equipo de medición AREXX
- Equipo de medición Gossen Metrawatt
- Equipo de medición PCE Instruments
- Equipo de medición CEM
- Equipo de medición SHX
- Equipo de medición Hameg
- Equipo de medición TDE Instruments
- Equipo de medición Joy-it
- Equipo de medición Circutor
- Equipo de medición Wachendorff
- Equipo de medición Klein Tools
- Equipo de medición Duro Pro
- Equipo de medición Aerospace Logic
- Equipo de medición GMW
- Equipo de medición Janitza
- Equipo de medición Sanwa
- Equipo de medición MSW
- Equipo de medición Weidmüller
- Equipo de medición Qubino
- Equipo de medición HT Instruments
- Equipo de medición P3 International
- Equipo de medición Beckmann & Egle
- Equipo de medición Seek Thermal
- Equipo de medición INFICON
- Equipo de medición Motrona
- Equipo de medición Tempo
- Equipo de medición Camille Bauer
- Equipo de medición Blebox
- Equipo de medición Seaward
- Equipo de medición HOZO Design
- Equipo de medición Corentium
- Equipo de medición GQ
- Equipo de medición Tektronix
- Equipo de medición Dostmann Electronic
- Equipo de medición Murideo
- Equipo de medición SecuTech
- Equipo de medición NetPeppers
- Equipo de medición SensoProtect
- Equipo de medición Water-i.d.
- Equipo de medición ZKETECH
- Equipo de medición LabNation
- Equipo de medición Electronics International
- Equipo de medición Cliff
- Equipo de medición ClimeMET
- Equipo de medición Fixpoint
- Equipo de medición Kübler
- Equipo de medición AkYtec
- Equipo de medición Stahlwille
- Equipo de medición MASIMO
- Equipo de medición Milesight
- Equipo de medición Enda
- Equipo de medición CSL
- Equipo de medición Ermenrich
- Equipo de medición PICO
- Equipo de medición Siglent
- Equipo de medición Noyafa
- Equipo de medición Stamos
Últimos Equipo de medición Manuales
26 Octubre 2024
25 Octubre 2024
25 Octubre 2024
24 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024
23 Octubre 2024