Gossen Metrawatt MAVOMAX 60 Manual de Usario

Lee a continuación 📖 el manual en español para Gossen Metrawatt MAVOMAX 60 (4 páginas) en la categoría Equipo de medición. Esta guía fue útil para 3 personas y fue valorada con 4.5 estrellas en promedio por 2 usuarios

Página 1/4
Operating Instructions
MAVOMAX 60, RK1, RK2/RK5
Interior Light Monitor 4/08.17
The MAVOMAX room light monitor is used wherever a constant illuminance has to be ensured. This applies, in
particular, to compliance with the room classes according to DIN 6868-157 of diagnostic rooms, examination
rooms with immediate diagnostics, dental diagnostic workstations and in teleradiology, where diagnostic
monitors and light boxes for medical applications are used. For universally usable rooms with dimmable lighting,
it supports the setting of the permissible illuminance for diagnostic.
During the transitional period for the quality assurance of image reproduction devices according to DIN V 6868-
57, the MAVOMAX can be used in order to extend the interval of constancy testing of image display devices for
veiling luminance maximum contrast to 6 months. In the case of repeat measurements within the framework and
of acceptance or constancy testing, veiling luminance need not be measured again and the 60-minute waiting
time until the image display device has reached a stable state is eliminated.
Initial Start- Up
The MAVOMAX is attached to the front edge of the monitor with the help of the double-faced tape at the bottom.
The front of the MAVO must be ush mounted to the front of the monitor. Please note that the MAVOMAX MAX
can no longer be removed after attaching it to your monitor. Forcible removal of the monitoring unit will
destroy the base plate of the MAVOMAX!
The MAVOMAX is supplied with electrical power by connecting it to either an available USB port or the included
mains power pack via the permanently attached USB cable. The MAVO is designed for continuous MAX
operation and is now ready for use.
The indicates that ambient light is within the permissible range for the interpretation of ndings. The green LED
red LED indicates that the brightness is not within the permissible range.
Technical Data
Monitoring range: MAVO MAX 60 (20…60 lux), RK1 ( 50 lux), 0… RK2/RK5 (50…100 lux)
Power supply: Via permanently attached USB cable, approx. 2.75 m long
Mains power pack: 100 … 240 V, 50/60 Hz, 0.15 A, for indoor use only
Dimensions: 40 x 33 x 23 mm
Protection class: Class II per VDE 0106, part 1
Maintenance
The instrument does not require any special maintenance if used in accordance with the operating instructions.
If the instrument becomes contaminated during use, clean the surface of the housing with a slightly moistened
cloth. Avoid the use of cleansers, abrasives or solvents. If at any time your MAVOMAX does not function to your
full satisfaction, please contact our service department.
Printed in Germany • Subject to change without notice
GOSSEN Foto- und Lichtmesstechnik GmbH I Lina-Ammon-Str.22 I D-90471 Nuremberg I Germany
Phone: +49 911 8602-181 I Fax: +49 911 8602-142 I e-mail: info@gossen-photo.de I www.gossen-photo.de

Especificaciones del producto

Marca: Gossen Metrawatt
Categoría: Equipo de medición
Modelo: MAVOMAX 60

¿Necesitas ayuda?

Si necesitas ayuda con Gossen Metrawatt MAVOMAX 60 haz una pregunta a continuación y otros usuarios te responderán




Equipo de medición Gossen Metrawatt Manuales

Equipo de medición Manuales

Últimos Equipo de medición Manuales